Undersampled Cine 3D tagging for rapid assessment of cardiac motion

Christian T Stoeck1, Robert Manka1,2, Peter Boesiger1 and Sebastian Kozerke1,3*

Author Affiliations

1 Institute for Biomedical Engineering, University and ETH Zurich, Gloriastrasse 35, 8092, Zurich, Switzerland

2 Department of Cardiology, University Hospital, Zurich, Switzerland

3 Imaging Sciences and Biomedical Engineering, King's College London, London, United Kingdom

For all author emails, please log on.

Journal of Cardiovascular Magnetic Resonance 2012, 14:60 doi:10.1186/1532-429X-14-60

Published: 30 August 2012

Abstract

Background

CMR allows investigating cardiac contraction, rotation and torsion non-invasively by the use of tagging sequences. Three-dimensional tagging has been proposed to cover the whole-heart but data acquisition requires three consecutive breath holds and hence demands considerable patient cooperation. In this study we have implemented and studied k-t undersampled cine 3D tagging in conjunction with k-t PCA reconstruction to potentially permit for single breath-hold acquisitions.

Methods

The performance of undersampled cine 3D tagging was investigated using computer simulations and in-vivo measurements in 8 healthy subjects and 5 patients with myocardial infarction. Fully sampled data was obtained and compared to retrospectively and prospectively undersampled acquisitions. Fully sampled data was acquired in three consecutive breath holds. Prospectively undersampled data was obtained within a single breath hold. Based on harmonic phase (HARP) analysis, circumferential shortening, rotation and torsion were compared between fully sampled and undersampled data using Bland-Altman and linear regression analysis.

Results

In computer simulations, the error for circumferential shortening was 2.8 ± 2.3% and 2.7 ± 2.1% for undersampling rates of R = 3 and 4 respectively. Errors in ventricular rotation were 2.5 ± 1.9% and 3.0 ± 2.2% for R = 3 and 4. Comparison of results from fully sampled in-vivo data acquired with prospectively undersampled acquisitions showed a mean difference in circumferential shortening of −0.14 ± 5.18% and 0.71 ± 6.16% for R = 3 and 4. The mean differences in rotation were 0.44 ± 1.8° and 0.73 ± 1.67° for R = 3 and 4, respectively. In patients peak, circumferential shortening was significantly reduced (p < 0.002 for all patients) in regions with late gadolinium enhancement.

Conclusion

Undersampled cine 3D tagging enables significant reduction in scan time of whole-heart tagging and facilitates quantification of shortening, rotation and torsion of the left ventricle without adding significant errors compared to previous 3D tagging approaches.

Keywords:
Motion analysis; Myocardial tagging; k-t undersampling; Principal component analysis; Image reconstruction