Skip to main content
  • Poster presentation
  • Open access
  • Published:

Low latency iterative reconstruction of first pass stress cardiac perfusion with physiological stress using graphical processing unit

Background

Cardiac MR perfusion has been shown to provide high diagnostic accuracy in detection of the coronary artery disease [1]. We have recently installed an MR-compatible supine bicycle mounted on the scanner table, which allows performing CMR perfusion immediately after physiologic stress. However, patients are unable to sustain a breathold after physical exercise, limiting the choice of acceleration techniques such as k-t approaches. Additionally, due to subject motion during exercise, coil sensitivity map are inaccurate resulting in imaging artifacts in conventional parallel imaging reconstruction. Compressed sensing (CS) is an alternative acceleration technique that enables high acceleration even without exploiting temporal dimension or need for coil maps. However, iterative CS reconstruction of randomly undersampled k-space is lengthy, performed off-line and is not usually integrated into the workflow of a clinical scan which requires viewing and initial assessment on the scanner console and storing the clinical images on the hospital PACS system. In this proposal, we aim to develop an accelerated iterative CS reconstruction workflow for reconstruction of CS acquired perfusion data using physical stress perfusion.

Methods

Figure 1 shows the workflow of the accelerated CMR perfusion reconstruction. After completion of the CMR perfusion sequence, the reconstruction process is manually started by CMR technologist using an in-house graphical user interface. All the subsequent reconstruction steps are then performed automatically without any user interaction. The raw data are pre-processed on the scanner workstation and sent to a dedicated computer for reconstruction (equipped with graphic processing unit (GPU) NVIDIA Tesla) and finally sent back to the scanner workstation and the PACS database. Pre- and post-processing are performed using the ReconFrame platform (Gyrotools, Zurich, Switzerland). The GPU-based CS reconstruction is implemented using a fast alternating minimization approach [2]. Since this reconstruction is iterative and voxel-independent for each iteration, the parallelization level of the GPU implementation was set to the voxel level. The presented workflow has been tested during bicycle ergometer stress CMR perfusion exams in healthy subjects using a 1.5T Philips scanner and a prospective 4× CS-accelerated CMR perfusion sequence. Computation time and latency of each reconstruction step was measured and compared to a non-parallelized implementation.

Figure 1
figure 1

Workflow of the iterative CS reconstruction for CMR perfusion immediately after supine bicycle physiologic stress using a remote workstation equipped with GPU. The undersampled k-space data are sent to the GPU workstation using hospital network. After reconstruction, the data are sent back to the scanner workstation for final post-processing using the ReconFrame platform for viewing and storage in clinical hospital PACS system and further analysis of images. This allows integration of CS-accelerated CMR perfusion into the clinical workflow with minimal user interaction.

Results

The proposed workflow allows reconstruction and viewing of the CS accelerated perfusion on the scanner console. The GPU-based implementation provides a 12-fold reduction in reconstruction time with an overall latency of 5 min 15s (Table 1).

Table 1 Computation time of the different reconstruction steps obtained for a complete CMR perfusion dataset.

Conclusions

GPU based CS-reconstruction significantly improved the reconstruction time and guarantee a minimal latency required for optimized clinical MR protocol for CMR perfusion during physical stress.

Funding

NIH:R01EB008743-01A2

References

  1. Greenwood : . The Lancet. 2012

    Google Scholar 

  2. Yang : . IEEE JSTSP. 2010

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Reprints and permissions

About this article

Cite this article

Roujol, S., Basha, T.A., Schülke, C. et al. Low latency iterative reconstruction of first pass stress cardiac perfusion with physiological stress using graphical processing unit. J Cardiovasc Magn Reson 15 (Suppl 1), E10 (2013). https://doi.org/10.1186/1532-429X-15-S1-E10

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/1532-429X-15-S1-E10

Keywords