Assessment of coronary artery disease using 3.0 T magnetic resonance coronary angiography: a national multicenter trial

Background

3.0T contrast enhanced whole-heart coronary magnetic resonance angiography (MRA) is a promising method for noninvasive, radiation-free detection and exclusion of obstructive coronary artery disease (CAD); however, the accuracy of this approach has not been determined in a multicenter trial.

Methods

An ECG-triggered, navigator-gated, inversion-recovery prepared, segmented gradient-echo sequence was used for image acquisition in 272 patients with suspected CAD at 8 hospitals. The accuracy of coronary MRA for detecting a 50% diameter reduction was determined using X-ray coronary angiography as the reference method. Using an intention-to-diagnose approach, all coronary arteries were included for the evaluation regardless of the image quality of coronary MRA to avoid overestimation of the diagnostic accuracy. Clinical Trial Registration—URL: http://clinicaltrials.gov. Unique identifier: NCT01024478.

Results

Acquisition of coronary MRA was successfully completed in 235 of 272 (86%) patients with average imaging time of 9.5±1.6 minutes. The areas under the receiver-operator characteristic curve from MRA images according to vessel- and patient-based analyses were 0.90 (95% confidence interval [CI]: 0.88 to 0.95) and 0.88 (95% CI: 0.83 to 0.93), respectively. The sensitivity and specificity of MRA on per-patient basis were 91% and 80%, respectively.

Figure 1
figure 1

Curved planar reconstruction (CPR) image (A), Sliding thin slab maximum intensity projection (MIP) image (B), MIP image of coronary tree (C), and volume-rendered image (D) detect coronary artery stenoses in the LAD (arrow) and first diagonal branch (arrowhead). Good agreement is observed between coronary MRA and X-ray coronary angiography.

Conclusions

Among patients who were scheduled to obtain conventional x-ray coronary angiography, we found that coronary MRA at 3.0T demonstrates high accuracy for detection of significant coronary artery stenosis. It warrants greater consideration as a suitable noninvasive method to exclude obstructive CAD.

Funding

National Basic Research Program 973 (grant no. 2010CB732600) from Ministry of Science and Technology, China; National Natural Science Foundation of China, grant number 30900355; National Institute of Health, grants numbers NIBIB EB002623 and NHLBI HL38698.

Author information

Authors and Affiliations

Authors

Rights and permissions

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Reprints and permissions

About this article

Cite this article

Yang, Q., Li, K., Sun, B. et al. Assessment of coronary artery disease using 3.0 T magnetic resonance coronary angiography: a national multicenter trial. J Cardiovasc Magn Reson 15 (Suppl 1), E5 (2013). https://doi.org/10.1186/1532-429X-15-S1-E5

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/1532-429X-15-S1-E5

Keywords