High-resolution modified look-locker inversion recovery (HR-MOLLI) for RV extracellular volume fraction at 3T and 1.5T: A feasibility study

Background

Cardiac MR (CMR) is the reference standard for assessing macroscopic myocardial scar. Determination of the extracellular volume fraction (ECV) by T1 estimation using the modified Look-Locker inversion recovery (MOLLI) correction enables quantitation of diffuse fibrosis. The purpose of this study is to evaluate the feasibility and reproducibility of an optimized high-resolution MOLLI (HR-MOLLI) technique at 3T and 1.5T for RV ECV calculation in healthy volunteers.

Methods

25 healthy volunteers (16 males, 41 ± 14.3 yrs) were scanned at 3T (MAGNETOM Skyra, Siemens AG, Healthcare Sector, Erlangen, Germany) and 15 (11 males, 46.4 ± 13.5 yrs) were scanned at 1.5T (MAGNETOM Aera). T1 mapping was performed in the axial orientation using an investigational HR-MOLLI technique with a 1 × 1 mm 2 in-plane resolution that applies motion correction with synthetic image estimation. Motion corrected images were used to generate parametric maps with (T1) and without (T1*) the MOLLI correction. The MOLLI sequence uses a 5 heart-beat (HB) acquisition, 3 HB recovery, 3 HB acquisition scheme with a single shot bSSFP diastolic readout. Images were acquired pre- and 10-25 minutes post- administration of 0.2 mmol/kg gadobenate dimeglumine (Multihance, Bracco Diagnostics, Monroe, NJ). T1 and T1* parametric maps were used to quantify the T1 of tissue and blood, respectively. Two reviewers quantified basal and mid RV free wall, interventricular septal, and lateral LV wall T1 values on T1 parametric maps. RV and LV ECV ranges were calculated assuming normal hematocrit values (women: 0.38-0.46, men: 0.42-0.54). Global ventricular ECV values were compared using the student's t-test. Intra and inter-observer variance was measured by the intraclass correlation coefficient (ICC).

Results

Representative T1 and T1* parametric maps are shown in Figure 1. One volunteer (3T) was excluded due to artifact. Table 1 shows RV and LV global ECV ranges by field strength and T1 BP estimation method. Global RV and LV ECV ranges were significantly different at 3T and 1.5T (p < 0.001). A significant influence of field strength was also noted for RV and LV ECV values (p < 0.05). Intraobserver (interobserver) variance for global RV and LV ECV was 0.78 and 0.92 (0.75 and 0.71) at 3T and 0.83 and 0.79 (0.58 and 0.71) at 1.5T, respectively.

Figure 1
figure 1

First row: 3T images Second row: 1.5T images Pre contrast T1, pre contrast T1*, post contrast T1, post contrast T1* from left to right for both rows.

Table 1 Global RV and LV ECV ranges at 3T and 1.5T as determined using T1 BP values from either the RV or LV on T1 and T1* parametric maps.

Conclusions

This feasibility study demonstrates that HR-MOLLI can quantitate the global RV ECV fraction at both 1.5 and 3T with good intra- and interobserver variance. Our results demonstrated a field-strength influence on RV and LV ECV values, and showed that ECV calculations using blood pool extimates without a look-locker correction are necessary at 3T.

Funding

Funding pending.

Author information

Authors and Affiliations

Authors

Rights and permissions

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Cite this article

Semaan, E., Spottiswoode, B.S., Freed, B. et al. High-resolution modified look-locker inversion recovery (HR-MOLLI) for RV extracellular volume fraction at 3T and 1.5T: A feasibility study. J Cardiovasc Magn Reson 16 (Suppl 1), P159 (2014). https://doi.org/10.1186/1532-429X-16-S1-P159

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/1532-429X-16-S1-P159

Keywords