Rapid phase contrast MRI with minimum time gradient waveform design using convex optimization
Journal of Cardiovascular Magnetic Resonance volume 16, Article number: W7 (2014)
Background
Phase contrast MRI (PC-MRI) sequences conventionally use triangular and trapezoidal gradient waveforms to construct flow compensated and flow encoded (FCFE) [1] or Bipolar [2] velocity encoding gradients. This approach does not make optimal use of the available gradient hardware for all time points, but convex optimization (CVX) can be used to design minimum-duration, arbitrarily shaped gradient waveforms subject to gradient hardware and pulse sequence constraints (e.g. target gradient moments and imaging parameters) [3]. Our objective was to quantitatively evaluate CVX PC-MRI measurement accuracy and sequence efficiency.
Methods
CVX was used for all gradient waveforms during Interval-1 (end of the slice-select gradient plateau to the beginning of the readout gradient plateau) and Interval-2 (end of the readout gradient plateau to the end of the TR, Figure 1), subject to (Gx2+Gy2+Gz2)0.5≤38 mT/m and (SRx2+SRy2+SRz2)0.5≤170 mT/m/ms. CVX PC-MRI also used asymmetric velocity encoding [4] for 2D through-plane velocity encoding to further improve sequence efficiency. Based on our recent evaluation of gradient spoiling in PC-MRI [6], the CVX sequence used 4π gradient spoiling while the FCFE sequence used a vendor hard-coded 9π gradient spoiling. A computer controlled flow phantom (sine wave with peak velocity of 150 cm/s) was used to validate that the arbitrarily shaped gradients used in CVX did not lead to measurement errors compared to FCFE. Bland-Altman analysis (reported as bias [lower 95%-CI, upper 95%-CI]) was used to compare FCFE and CVX velocity measurements with identical scan parameters (TE = 2.68 ms for 1.8 mm × 1.8 mm and 38 ms spatiotemporal resolution). Over a wide range of VENCs the minimum achievable TEMIN and TRMIN were compared for FCFE, Bipolar, and CVX using an otherwise fixed protocol. Sequence efficiency (E), defined as the ratio of the readout duration to the TR [5], was calculated for each VENC and sequence. Relative sequence efficiencies were compared as [ECVX-EFCFE]/ECVX and [ECVX-EBipolar]/ECVX.
Results
The Bland-Altman bias was 0.28 cm/s [-7.14 cm/s, 7.69 cm/s]. CVX produces the shortest TEMIN and TRMIN for every VENC (Figure 2). This leads to an increase in relative sequence efficiency by an average of 20 ± 6% (15% [min], 44% [max]) compared to FCFE and 24 ± 3% (20%, 31%) compared to Bipolar.
Conclusions
The Bland-Altman results indicate excellent agreement between CVX and FCFE (very low measurement bias and narrow 95%-CIs). Calculations show that ECVX can be increased by as much as 44%. Improvements in ECVX can be used to improve spatiotemporal resolution for a fixed breath hold duration or to reduce breath hold duration for an otherwise fixed protocol.
Funding
This work was enabled by research support from Siemens Medical Solutions and the Department of Radiological Sciences to DBE.
References
van Dijk P: JCAT. 1984
Bryant DJ: JCAT. 1984
Hargreaves BA: MRM. 2004
Bernstein MA: JMRI. 1992
Reeder SB: MRM. 1994
Zou Y: ISMRM. 2013
Author information
Authors and Affiliations
Rights and permissions
Open Access This article is published under license to BioMed Central Ltd. This is an Open Access article is distributed under the terms of the Creative Commons Attribution License ( https://creativecommons.org/licenses/by/2.0 ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. The Creative Commons Public Domain Dedication waiver ( https://creativecommons.org/publicdomain/zero/1.0/ ) applies to the data made available in this article, unless otherwise stated.
About this article
Cite this article
Middione, M.J., Wu, H.H. & Ennis, D.B. Rapid phase contrast MRI with minimum time gradient waveform design using convex optimization. J Cardiovasc Magn Reson 16 (Suppl 1), W7 (2014). https://doi.org/10.1186/1532-429X-16-S1-W7
Published:
DOI: https://doi.org/10.1186/1532-429X-16-S1-W7