Compressed sensing to accelerate magnetic resonance spectroscopic imaging: evaluation and application to 23Na-imaging of mouse hearts

Abstract

Background

Magnetic Resonance Spectroscopic Imaging (MRSI) has wide applicability for non-invasive biochemical assessment in clinical and pre-clinical applications but suffers from long scan times. Compressed sensing (CS) has been successfully applied to clinical 1H MRSI, however a detailed evaluation of CS for conventional chemical shift imaging is lacking. Here we evaluate the performance of CS accelerated MRSI, and specifically apply it to accelerate 23Na-MRSI on mouse hearts in vivo at 9.4 T.

Methods

Synthetic phantom data representing a simplified section across a mouse thorax were used to evaluate the fidelity of the CS reconstruction for varying levels of under-sampling, resolution and signal-to-noise ratios (SNR). The amplitude of signals arising from within a compartment, and signal contamination arising from outside the compartment relative to noise-free Fourier-transformed (FT) data were determined. Simulation results were subsequently verified experimentally in phantoms and in three mouse hearts in vivo.

Results

CS reconstructed MRSI data are scaled linearly relative to absolute signal intensities from the fully-sampled FT reconstructed case (R2 > 0.8, p-value < 0.001). Higher acceleration factors resulted in a denoising of the reconstructed spectra, but also in an increased blurring of compartment boundaries, particularly at lower spatial resolutions. Increasing resolution and SNR decreased cross-compartment contamination and yielded signal amplitudes closer to the FT data. Proof-of-concept high-resolution, 3-fold accelerated 23Na-amplitude maps of murine myocardium could be obtained within ~23 mins.

Conclusions

Relative signal amplitudes (i.e. metabolite ratios) and absolute quantification of metabolite concentrations can be accurately determined with up to 5-fold under-sampled, CS-reconstructed MRSI. Although this work focused on murine cardiac 23Na-MRSI, the results are equally applicable to other nuclei and tissues (e.g. 1H MRSI in brain). Significant reduction in MRSI scan time will reduce the burden on the subject, increase scanner throughput, and may open new avenues for (pre-) clinical metabolic studies.

Background

Magnetic Resonance Spectroscopic Imaging (MRSI) allows non-invasive investigation of regional metabolic processes in vivo, but suffers from long scan-times due to low metabolite concentrations, slow spatial encoding schemes, and low MR sensitivity (for nuclei other than protons). This versatile technique would benefit from a reduction in scan-time in order to make it more clinically applicable. Single- or multi-shot echo-planar spectroscopic imaging (EPSI) sequences [1-6] have been proposed to reduce MRSI scan-time. Indeed, Furuyama et al. have evaluated the performance of CS EPSI based 2D J-resolved spectroscopy [7,8]. However, EPSI is known to provide lower sensitivity compared to classical chemical shift imaging (CSI) [7]. Low metabolite concentrations and reduced MR sensitivity limit the applicability of parallel imaging. Conversely, Compressed Sensing (CS) is a technique for accelerating the inherently slow data acquisition process, and is well suited for MRSI due to its intrinsic denoising effect [9]. CS has been used to accelerate 1H- [10,11], hyperpolarized 13C- [12-14],23Na- [15], 31P-MRSI [16], and multi-dimensional NMR experiments [17,18]. The scan time reductions ranged from 2- [15,13] to 18-fold [14]. The application of CS to (pre-) clinical in vivo imaging necessitates thorough examination of the conditions under which it is robust. However, little systematic evaluation of the influence of SNR and spatial resolution on the achievable scan-time reductions and quantification of metabolite signals has been performed to date [7,13,19]. Whilst the work by Geethanath et al. [11] made an important step forward in the use of CS-MRSI, it generated some debate regarding the reliability of its use in the clinic [20,21]. We present a detailed performance evaluation of the CS-reconstruction developed by Geethanath et al. in the context of 1H-MRSI [11]. For the first time we then apply CS to 23Na-MRSI on mouse hearts demonstrating the potential of CS for high resolution spectroscopic imaging in vivo.

Methods

Synthetic phantom data

An in-silico noise-free phantom containing five compartments and representing a simplified mouse thorax was constructed in the image domain (256 × 256 voxels). Each compartment contained a single Lorentzian resonance at a unique frequency (Fig. 1a,b). This enabled quantification of inter-compartment signal contamination arising from CS reconstruction as the spatial origin of the signal could be determined from its frequency. These data were Fourier transformed (FT) into the time domain and the central N × N k-space points (N = 16, 24, 32, 48, 64) taken as low-resolution representations. Gaussian noise was added to vary the SNR in the image domain (2, 4, 8, 16, 32, 64, 128, 256 and ∞ (i.e. noise-free)), where SNR was defined as the ratio of peak-height to the standard deviation of the noise of the real part of the spectrum.

Fig. 1
figure 1

Layout of phantoms. a Layout of the virtual phantom. Five compartments were defined within the phantom representing aorta, myocardium, skeletal muscle, liver, and left ventricular blood pool (ROI1-5 respectively) with relative amplitudes of 5, 3, 1, 2, and 4. b The spectra corresponding to voxels within each compartment contain a single Lorentzian resonance at a unique frequency. Example spectra from each compartment within the virtual phantom, reconstructed at 64 × 64 voxels, are shown. Resonance frequencies for ROI1-5 were −1000, −500, 0, 500, 1000 Hz, respectively, with a T2 * of 50 ms. All spectra are plotted using the same axes. The small peak visible at −1000 Hz in ROI2 corresponds to signal contamination from ROI5. c 1H-MR image of the experimental phantom containing NaCl solutions (inner compartment: 20 mM NaCl; outer compartment: 100 mM NaCl). The two coaxial compartments on the left and right of the phantom contained 200 and 500 mM NaCl, respectively, and served as concentration references. The area used for data normalization is indicated by the dashed line (3.3 x 3.5 mm ROI), and the position of the surface coil used for 23Na signal reception is shown

Hardware

All 23Na-MRSI data were acquired on a horizontal 9.4 T MR system equipped with a VnmrS Direct Drive2 console and 60 mm i.d. 1 T/m imaging gradients (Agilent Technologies, USA). An actively decoupled 23Na quadrature birdcage transmit resonator (i.d. 39 mm) was used in combination with a 14 mm square surface receive coil (Rapid Biomedical, Germany). 1H anatomical images were acquired using a 39 mm quadrature birdcage resonator (Rapid Biomedical, Germany).

Phantom experiments

Validation experiments were conducted on a phantom comprised of two coaxial compartments with the inner compartment (i.d. 3.6 mm, volume ~ 1 ml) containing 20 mM NaCl, and the outer compartment (i.d. 13.5 mm, volume 15 ml) containing 100 mM NaCl (Fig. 1c). The phantom also contained two coaxial concentration references containing 200 and 500 mM NaCl, which, however, were not used for this study. Fully sampled, single average, 2D axial slice-selective MRSI data (60° flip angle, 30 × 30 mm FOV, 2 mm slice, 96 × 96 PE steps, TR/TE = 150/0.50 ms, 9216 FIDs, acquisition time ~24 mins) were acquired and repeated 32 times (total acquisition time: 12.8 h). The central N × N k-space points (N = 16, 24, 32, 48, 64, 96) from multiple repetitions were combined to give MRSI data sets with varying spatial resolution and SNR.

In vivo experiments

Cardiac 23Na-MRSI experiments were conducted on three female wild-type C57BL/6 mice (body weight 27.5 ± 4.0 g). After induction of anaesthesia (4 % isoflurane in 100 % oxygen), mice were positioned prone in a dedicated animal cradle with the heart centred over the surface coil, and maintained at 37 °C and 1.5–2.0 % isoflurane in 2 L/min oxygen flow throughout the MRSI experiments. Cardiac and respiratory signals were continuously monitored using an in-house developed ECG and respiratory gating device [22]. All investigations were approved by the local ethical review committee and conformed to the Animals (Scientific Procedures) Act 1986 (UK) incorporating European directive 2010/63/EU.

Using the same MRSI sequence as for the phantom data, threefold prospectively under-sampled 23Na data were acquired in mouse hearts in vivo in short-axis orientation (FOV 30 × 30 mm, 64 × 64 PE steps, 2 mm slice thickness, 8 averages, 40° flip angle, cardiac triggered, TR/TE = ~125/0.85 ms (i.e. one cardiac cycle), total acquisition time ~23 mins). Under-sampling was carried out as described previously [11].

Data analysis

All data were reconstructed and analysed in Matlab2013a (Mathworks, USA). Synthetic and acquired phantom data were retrospectively 1- to 5-fold under-sampled using a pseudo-random sampling scheme and reconstructed using a CS reconstruction described previously [11]. Signal amplitudes (i.e. peak maximum of the real component of the complex spectra), and contamination for CS reconstructed data and fully-sampled, FT reconstructed, data were compared. Signal amplitudes obtained from the CS reconstructed data were subject to linear regression analysis with the FT reconstructed fully-sampled data. A p-value of < 0.05 indicated a statistically significant correlation between the datasets. Signal contamination, defined as the total signal amplitude from resonances originating from outside the ROI detected within the ROI divided by the number of voxels in the ROI, was assessed for the synthetic phantom only. The 23Na myocardium/blood-ratio was estimated in vivo by calculating the mean ratio from six pairs of adjacent voxels located in the myocardium and the left ventricular blood-pool. Ratios are quoted as mean ± standard deviation. SNR was calculated for the myocardium by dividing the amplitude of the 23Na resonance in the spectrum by the standard deviation of the real component of the noise between 10–20 ppm, distant from the 23Na resonance; the chemical shift of the 23Na resonance was defined as 0 ppm.

Results

Typical CS reconstructed data from the virtual phantom (SNR = 32, 32 PE steps) are shown in Fig. 2, demonstrating similar spatial and relative signal intensity distribution to the FT data. Consistent with previous studies [13], low SNR (i.e. ≤ 8) resulted in an exaggerated noise floor in the CS reconstructed data preventing the reconstruction from converging; these data were excluded from further analysis. Although absolute signal intensities relative to the fully-sampled noise-free FT reconstructed case were not preserved, CS reconstruction resulted in a linear scaling of the data (right column in Fig. 2). Importantly, all CS reconstructed data demonstrated a high correlation with FT reconstructed noise-free fully-sampled data, with gradients ranging from 1.30 (R = 4, PE = 16, SNR = 16) to 5.73 (R = 1, PE = 64, SNR = 32), R2 values from 0.80 (R = 4, PE = 16, SNR = 16) to 0.98 (R = 1, PE = 64, SNR = ∞), and p-value always <0.001.

Fig. 2
figure 2

Example reconstruction of virtual phantom. Normalized signal amplitude maps (PE = 32, SNR = 32) are shown in the left column for (a) FT reconstructed fully-sampled data, and CS reconstructed data for (b) R = 1, (c) R = 3, and (d) R = 5. Example spectra from ROI2 and ROI3 are shown in the middle two columns. The ROI from which each resonance originates is indicated above the spectra. Due to blurring of compartment boundaries, there is signal contamination from compartments 3 and 5 evident in the spectrum for compartment 2; likewise the signal amplitude of the peaks in compartment 2 is reduced relative to the FT data. Note the denoising effect of the CS reconstruction as acceleration factor increases. All images and all spectra are plotted on the same scale. Correlation plots of FT reconstructed noise-free signal amplitudes against raw CS reconstructed amplitudes (i.e. not normalized) for the same data are depicted in the right column for R = 1, 3 and 5. Linear regression yielded slopes of 2.52, 2.61 and 2.53, with R2 values of 0.96, 0.96, and 0.94, respectively. All correlation plots share the same horizontal axis

Higher acceleration factor resulted in denoising of the reconstructed spectra, but also in increased blurring of compartment boundaries, particularly at lower resolutions. This is qualitatively illustrated by the boundary of ROI3 (Fig. 3). While the FT reconstruction of the fully sampled data yielded the sharpest transition from signal within ROI3 to the outside, oscillations in signal amplitude outside the compartment are visible as a result of the point-spread-function of the uniformly weighted acquisition. Conversely, CS reconstructions of (R = 1 or 3) reduced these oscillations at the boundary transition, but resulted in an increased near-constant signal-floor outside the compartment, which decreased with improving spatial resolution.

Fig. 3
figure 3

Effect on spatial resolution of CS reconstruction. Mean signal profiles at the boundary of compartment 3 of the virtual phantom as an illustration of the point-spread function using (a) 16, (b) 32, and (c) 64 phase encoding steps are shown for FT reconstruction (bold black line) and CS reconstructions with R = 1, 3 and 5 (red, purple, and blue lines respectively). The profile from equivalent acquisition-weighted (Hanning) FT reconstructed MRSI data is included for comparison (dotted line). All data are normalized to a maximum amplitude of 1.0

Figure 4 shows the color-coded (a,b; top row) signal amplitude and (c,d; bottom row) contamination for ROI2-4 of the virtual phantom, normalized to the mean signal intensity from ROI1. Data were normalized to the signal from ROI1 as this compartment provided a defined signal and had dimensions similar to those that would be expected from a concentration reference phantom that could reasonably used in murine cardiac in vivo experiments. Data are presented as a function of spatial resolution, acceleration factor, and SNR for the CS- (a, c; left column) and the FT-reconstructions (b, d; right column). Unsurprisingly, the highest contamination, and correspondingly lower signal amplitudes were found for the CS reconstructed data of ROI2 (representing left-ventricular myocardium) at low resolution. Similarly, low amplitudes and high contamination were found for CS reconstructions of all ROIs at lowest SNR and increasing under-sampling factors. Increasing SNR and resolution resulted in CS reconstructions that more closely resembled the FT reconstruction of fully sampled data, while contamination decreased to levels lower than in those for the FT reconstruction.

Fig. 4
figure 4

Signal amplitude and contamination for virtual phantom reconstructions. Mean signal amplitude from within an ROI normalized to the mean signal of ROI1 is plotted against number of phase encoding steps (Res; PE = 16, 24, 32, 48, 64), ROI number (ROI; 2–5), acceleration factor (R; 1–5), and SNR (∞, 256, 128, 64, 32, and 16) for (a) CS and (b) FT reconstructions; plots share the same scale. The CS plot (a) is divided into 5×4 element rectangles, each corresponding to CS reconstructions at a given SNR/spatial resolution. Columns correspond to acceleration factor (R = 1-5; left to right) and rows to ROI2-5 (top to bottom). SNR decreases towards the right and resolution increases towards the bottom. The FT plot (b) is divided into 6×4 element rectangles, each rectangle corresponding to FT reconstructions carried out at a given spatial resolution; columns correspond to SNR (∞, 256, 128, 64, 32, and 16; left to right) and rows to ROI2-5 (top to bottom). The number of phase encoding steps increases towards the bottom of the plot. Triangles below/left of the plots indicate parameter magnitude with the thin end corresponding to the lowest value. If a CS reconstruction performs identically to the equivalent FT reconstruction, then the column of the CS plot (a) should appear identical to the corresponding column of the FT plot (b). Mean signal contamination arising from outside an ROI detected within that ROI normalized to the mean signal from ROI1 is plotted against number of phase encoding steps, ROI number, CS acceleration factor, and SNR for both (c) CS and (d) FT reconstructions; plots share the same scale. Interpretation of the plots is analogous to that of the amplitude plots above. Increasing SNR and decreasing acceleration factor improves agreement between CS and FT reconstructions. Note the separate scale bars for amplitude and contamination plots

In Fig. 5, sodium intensity maps from the compartmented phantom are shown for FT reconstructed fully-sampled, and CS-reconstructed data (R = 1, 3 and 5; 32 PE steps). The gradient in signal intensities along the vertical axis of the images reflects the receive profile of the surface coil. Horizontal and vertical profile, correlation plots (normalized FT vs. CS reconstructed data), and Bland-Altman analysis for the amplitudes confirmed the behaviour observed in the virtual phantom: CS reconstruction resulted in a high correlation between CS and FT reconstructed data, in signal denoising, and in increased blurring of compartment boundaries with increased under-sampling.

Fig. 5
figure 5

CS reconstruction of retrospectively under-sampled phantom data. Results of CS reconstruction of retrospectively under-sampled phantom CSI data acquired with one average and 32 × 32 PE steps. Images represent the signal amplitude (magnitude-modulated by the receive profile of the surface coil, the position of which is indicated in the top left panel) following FT reconstruction of fully-sampled data, and CS reconstruction of data with R = 1, 3, and 5 (top to bottom). All images are plotted on the same scale. The white dashed lines in the top left panel illustrate the position of the displayed profiles (second column from left); solid and dashed lines indicate vertical and horizontal profiles, respectively. Note the reduction in noise around the phantom arising from the CS reconstruction for R = 1 relative to the FT reconstruction and smoother profiles across the phantom. Increasing acceleration factor and decreasing SNR resulted in blurring of the boundaries of the phantom in the images. The third column shows the correlation between the normalized signal amplitude of fully-sampled FT reconstructed data (x-axis) and of the CS reconstructed under-sampled data (y-axis); data were normalized to the mean signal arising from the ROI indicated in Fig. 1c. Linear regression yielded slopes of 1.09, 0.98, and 0.88, with R2 values of 0.98, 0.96, and 0.89, respectively. The right-hand column shows Bland-Altman plots for normalized FT- and CS-reconstructed signal amplitudes. The mean difference in signal amplitude is indicated by a solid line, and 95 % confidence interval by dashed lines. The value of the mean difference in amplitude is indicated on the plots. All plot amplitudes are in arbitrary units

A prospectively 3-fold under-sampled, CS-reconstructed, myocardial 23Na-map in short-axis orientation obtained in vivo from a mouse overlaid on an anatomical 1H-image, and corresponding spectra, are presented in Fig. 6. Blood/myocardium amplitude ratios of 0.45 ± 0.07, 0.48 ± 0.09, and 0.47 ± 0.08 were obtained in vivo; assuming a blood 23Na concentration of ~79 mM [23], myocardial 23Na concentrations of 35.6, 37.9, and 37.1 mM can be estimated for the three experiments. The SNR in the anterior myocardium (close to the surface coil) was 42.1 ± 1.1, 49.2 ± 1.4, and 75.1 ± 1.8 (mean ± SD) whilst in the posterior myocardium (distant from the surface coil) it was 36.6 ± 1.3, 36.5 ± 2.4, and 61.2 ± 1.5 (mean ± SD), respectively.

Fig. 6
figure 6

Representative in vivo CS three-fold accelerated cardiac MRSI. CS reconstructed prospectively under-sampled 23Na CSI (R = 3) of a murine heart in short-axis orientation (colour) overlaid onto a 1H anatomical image (black and white). The position of the surface coil is indicated. The concentration reference phantom (CR) has been omitted from the CSI data. The difference in Na concentration between the myocardium and blood pool is clearly reflected in the color-coded signal amplitudes of the 23Na image. Example phased spectra for the myocardium (*) and left ventricular blood (+) are shown on the right. Both spectra are plotted on the same scale

Discussion and conclusions

MRS(I) provides a unique method for investigation of metabolic processes in living tissue and has been successfully applied in clinical studies of brain and cancer metabolism (e.g. [24-26]). Specifically, 23Na-MRSI has the potential to monitor myocardial ion homeostasis in models of cardiac disease. Widespread application of MRSI remains limited, especially in the clinic, due to significant technical challenges, low sensitivity and associated long scan-times. A reduction in acquisition time without sacrificing accuracy would represent an important step towards making MRSI more widely applicable. The CS algorithm used in this work has been applied previously to study brain and prostate tumours with 1H-MRSI [11], however our study offers several novel aspects. By providing a thorough validation of the reconstruction, we sought to determine the (pre-) clinical applicability of CS-accelerated MRSI. The use of synthetic data ensured that the results were generally applicable, independent of organ/tissue and nucleus. It also permitted an assessment of cross-compartment signal contamination, which is not possible with (noise-free model of) clinical data where compartments share similar metabolite populations. To the best of our knowledge, only one study has reported the application of 23Na-MRSI to murine hearts in vivo [27]. 23Na-MR spectroscopy commonly utilises only a single resonance arising from both intra- and extra-cellular sodium, which may therefore also be detected with MR imaging. However, we used this nucleus to prove the concept of the technique, with the ultimate aim of combining it with multiple-quantum filters to distinguish intra- from extra-cellular sodium. Thus, application of CS to 23Na-MRSI could significantly reduce scan-time making metabolic data acquisition possible within clinically acceptable times.

There are several technical challenges to be faced in applying CS to 23Na-MRSI of (murine) cardiac tissue: large sodium concentration differences between tissues, and thus MRSI signal intensity variations, lie in close proximity (e.g. blood vs. myocardium vs. lungs), which can be difficult to reconstruct using CS; the influence of SNR and spatial resolution on CS reconstruction fidelity had to be characterized.

Virtual phantom data, which mimicked the appearance and challenges of an in vivo cross-section through a murine thorax, was used to characterize CS performance. For ROI2 (left-ventricular myocardium) this resulted in a wall thickness of between 1 and 3 voxels corresponding to ~0.5-1.5 mm in typical in vivo mouse cardiac MRSI experiment. We found that absolute signal amplitudes following CS reconstruction were not preserved, but correlated linearly with those obtained in the (noise-free) conventional FT reconstruction (i.e. “ground truth”). This finding has two immediate implications. Firstly, linear scaling of the amplitude means relative amplitudes (i.e. metabolite ratios) remain unaffected by CS reconstruction. Consequently, absolute quantification of metabolite concentration relative to a signal from a compartment of known concentration remains accurate. Secondly, the calculation of root-mean-square errors, which are typically used as a measure of data fidelity, would yield high values and not reflect the quality of the reconstruction. Thus reconstructed signal amplitudes and signal contamination, normalized to the mean signal intensity from a defined ROI, were chosen as metrics for this purpose (Fig. 4). For the CS reconstructions, increasing spatial resolution reduced signal contamination (bottom row, Fig. 4) and blurring of compartment boundaries (Fig. 3). Low-resolution FT reconstructed data yielded signal oscillations inside and outside the compartment due to convolution of the image with the FT of a step function. CS reconstruction reduced these oscillations at the cost of near-constant signal level outside the compartment, which increased with higher under-sampling factors and/or decreasing resolution. Whilst there is a straightforward relationship between data sampling and point spread function for FT reconstruction, the relationship is more complex for CS. The CS reconstruction used here assumes that an image consists of regions of approximately homogeneous signal with relatively sharp discontinuities in signal intensity between regions, and that the resulting reconstruction must be consistent with the acquired k-space data. These assumptions give rise to some blurring of the boundaries between homogeneous regions resulting in an apparent loss of spatial resolution and out-of-voxel signal contamination. Some understanding of the origins of a signal can be inferred from the spatial distribution of the reconstructed signal across a sharp boundary, for example at the edge of a phantom where the signal transitions from that originating from within the phantom to the void surrounding the phantom. Signal present in the void surrounding the phantom indicates of out-of-voxel contamination present in that voxel; Fig. 3 demonstrates this effect of CS and reveals that the out-of-voxel contamination approaches that present in FT reconstructed data as spatial resolution increases. CS can result in low-level signal contamination arising from signals a considerable distance outside the target voxel.

Our findings suggest in agreement with the literature [11], that up to 5-fold under-sampling of (pre-)clinical MRSI will be possible without significant impact on data fidelity given modest SNR. This conclusion was subsequently confirmed in retrospectively under-sampled phantom experiments. Prospectively accelerated in vivo data were also acquired. Notably, our murine study was at the lower end of the SNR requirements due to the very high spatial resolution; Neuberger et al. report that a voxel volume of 1 mm3 could be achieved within ~90 mins [27], which was over twice the volume used in our study. Thus, a conservative under-sampling factor of 3 was chosen in vivo. Additionally, the reconstruction of the phantom and in vivo data demonstrated that CS accurately reproduces spatially inhomogeneous signals arising from the receive profile of the surface coil used to acquire the data. The quantified myocardium/blood amplitude ratios agreed well with previously reported values of 0.42 ± 0.15 for mouse [27] and 0.57 for humans [23]. A 3-fold reduction in scan time to ~23 mins for a single slice makes our protocol suitable for use as part of a multi-parametric MR-phenotyping investigation, or to study severely sick animals by reducing the anaesthetic burden. Assuming that the typical mouse heart is approximately 8 mm in length, full heart coverage using 2D CS-MRSI as described would therefore take ~92 mins; this represents half the voxel volume described by Neuberger et al. within the same scan time. Furthermore, use of through-plane sparsity information in the CS reconstruction should allow further acceleration of 2D multislice or 3D CS-MRSI without loss of reconstruction fidelity. Compressed sensing has undoubtedly a significant role to play in the future of preclinical and clinical MRSI. Increases in total tissue sodium concentration of between 150 and 330 % have been observed for a variety of pathologies [28-31]. Fig. 3 would suggest that for resolutions at or above 32 × 32 PE steps, the level of potential signal contamination in a voxel arising from CS reconstruction would have an approximate upper limit of 20 %. However, in the presence of this level of signal contamination, such pathologies should be detectable. In general, the SNR of the data will fundamentally limit the ability to detect pathological abnormalities; whilst FT performs better than CS at very low SNR, accurate data fitting/quantification under these conditions may prove limiting even for FT reconstruction. Whilst the experimental results presented here focus on MRSI and its application to 23Na, the simulation results apply equally well to 1H, 31P, or 13C MRSI. Prospectively under-sampling data can significantly reduce MRSI data acquisition to a clinically acceptable time, reducing the burden on the patient and increasing scanner throughput without sacrificing diagnostic accuracy.

Abbreviations

MRSI:

Magnetic resonance spectroscopic imaging

CS:

Compressed sensing

SNR:

Signal-to-noise ratio

FT:

Fourier-transform

EPSI:

Echo-planar spectroscopic imaging

CSI:

Chemical shift imaging

R:

Acceleration/undersampling factor

References

  1. Mansfield P. Spatial mapping of the chemical shift in NMR. Magn Reson Med. 1984;1(3):370–86.

    Article  CAS  PubMed  Google Scholar 

  2. Doyle M, Mansfield P. Chemical-shift imaging: a hybrid approach. Magn Reson Med. 1987;5:255–61.

    Article  CAS  PubMed  Google Scholar 

  3. Guilfoyle DN, Mansfield P. Chemical-shift imaging. Magn Reson Med. 1985;2:479–89.

    Article  CAS  PubMed  Google Scholar 

  4. Norris DG. Ultrafast low-angle RARE: U-FLARE. Magn Reson Med. 1991;17(2):539–42.

    Article  CAS  PubMed  Google Scholar 

  5. Jakob PM, Kober F, Pohmann R, Haase A. Single-shot spectroscopic imaging (SISSI) using a PEEP/BURST hybrid. J Magn Reson B. 1996;110(3):278–83.

    Article  CAS  PubMed  Google Scholar 

  6. Jakob PM, Ziegler A, Doran SJ, Decorps M. Echo-time-encoded burst imaging (EBI): a novel technique for spectroscopic imaging. Magn Reson Med. 1995;33(4):573–8.

    Article  CAS  PubMed  Google Scholar 

  7. Furuyama JK, Wilson NE, Burns BL, Nagarajan R, Margolis DJ, Thomas MA. Application of compressed sensing to multidimensional spectroscopic imaging in human prostate. Magn Reson Med. 2012;67(6):1499–505. doi:10.1002/mrm.24265.

    Article  PubMed  Google Scholar 

  8. Sarma MK, Nagarajan R, Macey PM, Kumar R, Villablanca JP, Furuyama J, et al. Accelerated echo-planar J-resolved spectroscopic imaging in the human brain using compressed sensing: a pilot validation in obstructive sleep apnea. AJNR Am J Neuroradiol. 2014;35(6 Suppl):S81–9. doi:10.3174/ajnr.A3846.

    Article  CAS  PubMed Central  PubMed  Google Scholar 

  9. Lustig M, Donoho D, Sparse PJM, MRI. The application of compressed sensing for rapid MR imaging. Magn Reson Med. 2007;58(6):1182–95. doi:10.1002/mrm.21391.

    Article  PubMed  Google Scholar 

  10. Cao P, Wu EX. Accelerating phase-encoded proton MR spectroscopic imaging by compressed sensing. J Magn Reson Imaging. 2014. doi:10.1002/jmri.24553.

    Google Scholar 

  11. Geethanath S, Baek HM, Ganji SK, Ding Y, Maher EA, Sims RD, et al. Compressive sensing could accelerate 1H MR metabolic imaging in the clinic. Radiology. 2012;262(3):985–94. doi:10.1148/radiol.11111098.

    Article  PubMed Central  PubMed  Google Scholar 

  12. Hu S, Lustig M, Balakrishnan A, Larson PE, Bok R, Kurhanewicz J, et al. 3D compressed sensing for highly accelerated hyperpolarized (13)C MRSI with in vivo applications to transgenic mouse models of cancer. Magn Reson Med. 2010;63(2):312–21. doi:10.1002/mrm.22233.

    Article  CAS  PubMed Central  PubMed  Google Scholar 

  13. Hu S, Lustig M, Chen AP, Crane J, Kerr A, Kelley DA, et al. Compressed sensing for resolution enhancement of hyperpolarized 13C flyback 3D-MRSI. J Magn Reson. 2008;192(2):258–64. doi:10.1016/j.jmr.2008.03.003.

    Article  CAS  PubMed Central  PubMed  Google Scholar 

  14. Larson PE, Hu S, Lustig M, Kerr AB, Nelson SJ, Kurhanewicz J, et al. Fast dynamic 3D MR spectroscopic imaging with compressed sensing and multiband excitation pulses for hyperpolarized 13C studies. Magn Reson Med. 2011;65(3):610–9. doi:10.1002/mrm.22650.

    Article  CAS  PubMed Central  PubMed  Google Scholar 

  15. Madelin G, Chang G, Otazo R, Jerschow A, Regatte RR. Compressed sensing sodium MRI of cartilage at 7 T: preliminary study. J Magn Reson. 2012;214(1):360–5. doi:10.1016/j.jmr.2011.12.005.

    Article  CAS  PubMed Central  PubMed  Google Scholar 

  16. Askin NC, Atis B, Ozturk-Isik E. Accelerated phosphorus magnetic resonance spectroscopic imaging using compressed sensing. Conf Proc IEEE Eng Med Biol Soc. 2012;2012:1106–9. doi:10.1109/EMBC.2012.6346128.

    PubMed  Google Scholar 

  17. Bostock MJ, Holland DJ, Nietlispach D. Compressed sensing reconstruction of undersampled 3D NOESY spectra: application to large membrane proteins. J Biomol NMR. 2012;54(1):15–32. doi:10.1007/s10858-012-9643-4.

    Article  CAS  PubMed  Google Scholar 

  18. Kazimierczuk K, Orekhov VY. A comparison of convex and non-convex compressed sensing applied to multidimensional NMR. J Magn Reson. 2012;223:1–10. doi:10.1016/j.jmr.2012.08.001.

    Article  CAS  PubMed  Google Scholar 

  19. Heikal AA, Wachowicz K, Fallone BG. MTF behavior of compressed sensing MR spectroscopic imaging. Med Phys. 2013;40(5):052302. doi:10.1118/1.4800642.

    Article  CAS  PubMed  Google Scholar 

  20. Barker PB, Ernst TM. Acceleration of 1H MR metabolic imaging with compressed sensing. Radiology. 2013;266(2):686. doi:10.1148/radiol.12121692.

    Article  PubMed  Google Scholar 

  21. Geethanath S, Kodibagkar VD. Response Radiology. 2013;266(2):686–7.

    Google Scholar 

  22. Cassidy PJ, Schneider JE, Grieve SM, Lygate C, Neubauer S, Clarke K. Assessment of motion gating strategies for mouse magnetic resonance at high magnetic fields. J Magn Reson Imaging. 2004;19(2):229–37.

    Article  PubMed  Google Scholar 

  23. Ouwerkerk R, Weiss RG, Bottomley PA. Measuring human cardiac tissue sodium concentrations using surface coils, adiabatic excitation, and twisted projection imaging with minimal T2 losses. J Magn Reson Imaging. 2005;21(5):546–55.

    Article  PubMed  Google Scholar 

  24. Kwock L, Smith JK, Castillo M, Ewend MG, Collichio F, Morris DE, et al. Clinical role of proton magnetic resonance spectroscopy in oncology: brain, breast, and prostate cancer. Lancet Oncol. 2006;7(10):859–68. doi:10.1016/S1470-2045(06)70905-6.

    Article  PubMed  Google Scholar 

  25. Glunde K, Bhujwalla ZM. Metabolic tumor imaging using magnetic resonance spectroscopy. Semin Oncol. 2011;38(1):26–41. doi:10.1053/j.seminoncol.2010.11.001.

    Article  PubMed Central  PubMed  Google Scholar 

  26. Tran T, Ross B, Lin A. Magnetic resonance spectroscopy in neurological diagnosis. Neurol Clin. 2009;27(1):21–60. doi:10.1016/j.ncl.2008.09.007. 1.

    Article  PubMed  Google Scholar 

  27. Neuberger T, Greiser A, Nahrendorf M, Jakob PM, Faber C, Webb AG. 23Na microscopy of the mouse heart in vivo using density-weighted chemical shift imaging. MAGMA. 2004;17(3–6):196–200.

    Article  CAS  PubMed  Google Scholar 

  28. Ouwerkerk R, Bleich KB, Gillen JS, Pomper MG, Bottomley PA. Tissue sodium concentration in human brain tumors as measured with 23Na MR imaging. Radiology. 2003;227(2):529–37. doi:10.1148/radiol.2272020483.

    Article  PubMed  Google Scholar 

  29. Constantinides CD, Weiss RG, Lee R, Bolar D, Bottomley PA. Restoration of low resolution metabolic images with a priori anatomic information: 23Na MRI in myocardial infarction. Magn Reson Imaging. 2000;18(4):461–71.

    Article  CAS  PubMed  Google Scholar 

  30. Wetterling F, Tabbert M, Junge S, Gallagher L, Macrae IM, Fagan AJ. A double-tuned (1)H/(23)Na dual resonator system for tissue sodium concentration measurements in the rat brain via Na-MRI. Phys Med Biol. 2010;55(24):7681–95. doi:10.1088/0031-9155/55/24/019.

    Article  PubMed  Google Scholar 

  31. Kim RJ, Judd RM, Chen EL, Fieno DS, Parrish TB, Lima JA. Relationship of elevated 23Na magnetic resonance image intensity to infarct size after acute reperfused myocardial infarction. Circulation. 1999;100:185–92.

    Article  CAS  PubMed  Google Scholar 

Download references

Acknowledgements

The authors would like to thank Ms Vicky Thornton for her help with the in vivo experiments. This project was funded by a MRC/EPSRC Grant (G0600829). JES is a British Heart Foundation Senior Basic Science Research Fellow (FS/11/50/29038), and the authors acknowledge the Wellcome Trust Core Award (grant 090532/Z/09/Z).

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Jürgen E. Schneider.

Additional information

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

MLM designed the study, conducted the simulations and MR experiments, carried out the data analysis, and drafted the manuscript. SG and VK wrote the compressed sensing algorithm. CAL was involved in the study design and in vivo experiments. JES was involved in the design of the study, data analysis and in drafting of the manuscript. All authors read and approved the final manuscript.

Rights and permissions

Open Access  This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made.

The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder.

To view a copy of this licence, visit https://creativecommons.org/licenses/by/4.0/.

The Creative Commons Public Domain Dedication waiver (https://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Maguire, M.L., Geethanath, S., Lygate, C.A. et al. Compressed sensing to accelerate magnetic resonance spectroscopic imaging: evaluation and application to 23Na-imaging of mouse hearts. J Cardiovasc Magn Reson 17, 45 (2015). https://doi.org/10.1186/s12968-015-0149-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12968-015-0149-6

Keywords