Saturation pulse design for quantitative myocardial T1 mapping

Abstract

Background

Quantitative saturation-recovery based T1 mapping sequences are less sensitive to systematic errors than the Modified Look-Locker Inversion recovery (MOLLI) technique but require high performance saturation pulses. We propose to optimize adiabatic and pulse train saturation pulses for quantitative T1 mapping to have <1 % absolute residual longitudinal magnetization (|MZ/M0|) over ranges of B0 and \( {\widehat{B}}_1 \) (B1 scale factor) inhomogeneity found at 1.5 T and 3 T.

Methods

Design parameters for an adiabatic BIR4-90 pulse were optimized for improved performance within 1.5 T B0 (±120 Hz) and \( {\widehat{B}}_1 \) (0.7–1.0) ranges. Flip angles in hard pulse trains of 3–6 pulses were optimized for 1.5 T and 3 T, with consideration of T1 values, field inhomogeneities (B0 = ±240 Hz and \( {\widehat{B}}_1 \)=0.4–1.2 at 3 T), and maximum achievable B1 field strength. Residual MZ/M0 was simulated and measured experimentally for current standard and optimized saturation pulses in phantoms and in-vivo human studies. T1 maps were acquired at 3 T in human subjects and a swine using a SAturation recovery single-SHot Acquisition (SASHA) technique with a standard 90°-90°-90° and an optimized 6-pulse train.

Results

Measured residual MZ/M0 in phantoms had excellent agreement with simulations over a wide range of B0 and \( {\widehat{B}}_1 \). The optimized BIR4-90 reduced the maximum residual |MZ/M0| to <1 %, a 5.8× reduction compared to a reference BIR4-90. An optimized 3-pulse train achieved a maximum residual |MZ/M0| <1 % for the 1.5 T optimization range compared to 11.3 % for a standard 90°-90°-90° pulse train, while a 6-pulse train met this target for the wider 3 T ranges of B0 and \( {\widehat{B}}_1 \). The 6-pulse train demonstrated more uniform saturation across both the myocardium and entire field of view than other saturation pulses in human studies. T1 maps were more spatially homogeneous with 6-pulse train SASHA than the reference 90°-90°-90° SASHA in both human and animal studies.

Conclusions

Adiabatic and pulse train saturation pulses optimized for different constraints found at 1.5 T and 3 T achieved <1 % residual |MZ/M0| in phantom experiments, enabling greater accuracy in quantitative saturation recovery T1 imaging.

Background

Saturation pulses are commonly used in MRI pulse sequences to prepare longitudinal magnetization to a known zero state, independent of previous acquisitions, and generate spin–lattice (T1) relaxation contrast. Several methods for T1 quantification in the heart are based on multiple image acquisitions with variable saturation recovery times such as the saturation recovery turbo flash (SRTFL) [1], short acquisition period (SAP-T1) [2], and the recently proposed SAturation recovery single-SHot Acquisition (SASHA) [3] sequences.

All of these techniques assume ideal saturation efficiency, and poor saturation performance results in errors in calculated T1 values. The saturation efficiency can be quantified using residual longitudinal magnetization (MZ/M0), which can range from −1 (full inversion) to 1 (no effect) and 0 for perfect saturation. In the standard three-parameter exponential model for SASHA T1 data, saturation efficiency is a fitted parameter, but imperfect saturation may still result in errors in the calculated T1 values due to residual magnetization remaining between heartbeats [3]. Two-parameter fitting of SASHA data assuming ideal saturation has been shown to substantially reduce variability of calculated T1 values [4], but results in larger T1 errors when there is imperfect saturation.

Composite saturation pulses consisting of trains of shaped RF pulses with numerically optimized flip angles have been designed for several different ranges of B0 (off-resonance) and \( {\widehat{B}}_1 \) scale factors (ratio of actual radiofrequency field strength to nominal radiofrequency field strength). For example, enhanced water suppression has been achieved over narrow B0 and \( {\widehat{B}}_1 \) ranges for spectroscopy applications at 1.5 T [5] and numerically optimized adiabatic pulses have been used for wide B0/\( {\widehat{B}}_1 \) ranges at 7 T [6]. Hard RF pulse trains have also been proposed [7] and a train of 3 pulses with flip angles tailored for improved performance over B0 and \( {\widehat{B}}_1 \) ranges expected at 3 T has been evaluated in-vivo [8]. However, the maximum residual MZ/M0 of >8 % of this design may be a significant source of error when used in quantitative imaging sequences.

Adiabatic BIR4-90 saturation pulses [9] and hybrid adiabatic-rectangular saturation pulses [10] have also been proposed, but the high B1 field strength required to meet the adiabatic criteria may exceed the allowable specific absorption rate (SAR), particularly on larger subjects. On higher B0 field strength magnets, lower \( {\widehat{B}}_1 \) scale factors may also reduce the effective B1 field to below the adiabatic limit and result in poor performance [8, 11].

In this work, we expand upon existing literature regarding the design of saturation pulse trains and adiabatic saturation pulses with the primary goal of significantly improving performance to less than 1 % residual |MZ/M0| over ranges of B0 and \( {\widehat{B}}_1 \) values found at 1.5 T and 3 T, given the practical constraints of peak B1 transmit field strength. The performance of the proposed saturation pulses are simulated, experimentally validated, and compared to current standard saturation pulses in phantom experiments and in-vivo human studies. The effect of saturation pulse performance on the two-parameter SASHA T1 mapping technique is also characterized through simulations and in-vivo studies on humans and a swine.

Methods

B1 field strength limitations

The maximum B1 field strength is predominately limited by the achievable RF amplifier voltage for a given RF coil. These B1 limitations reduce the off-resonance performance of both adiabatic pulses and hard pulse trains, imposing an additional constraint in their optimization. For the MRI systems used in this study, a reference voltage was calculated for each subject corresponding to the amplifier voltage required to achieve a B1 field strength of 11.7 μT, equivalent to a 180° flip angle with a 1 ms rectangular pulse. The maximum achievable B1 was calculated by multiplying 11.7 μT by the ratio of maximum voltage to the reference voltage. The maximum achievable B1 field strength was calculated in this way for subjects on Siemens MAGNETOM 1.5 T Aera and 3 T Skyra systems over an approximate period of 6 months, with information collected as quality assurance data and analysis approved by the NIH Office of Human Subject Research.

Spoiler gradient design

Spoiler gradients in between RF pulses in a saturation pulse train were designed to minimize transverse magnetization. In contrast to previous pulse train designs where inter-pulse spoilers were polarity cycled along a single direction [7, 8, 10], inter-pulse spoilers in the proposed design were cycled along multiple gradient directions in order to dephase magnetization in all spatial directions and reduce the likelihood that spoiler gradients could be unwound during image readouts. Spoiler durations and total areas were also varied to minimize the potential formation of coherent stimulated echoes, with no two spoilers on the same axis having the same area. The implementation in this study uses a gradient strength of 24 mT/m per channel, which is achievable on most modern MRI scanners. At this strength, the minimum spoiler area in the 6-pulse train was 60 mTms/m, equivalent to 2π phase dispersion across 0.4 mm, with other spoiler areas and directions detailed in Fig. 1a. Spoiler designs for shorter pulse trains with n pulses share the same design as the first n spoiler gradients in the 6 pulse design, as shown for a 4-pulse train in Fig. 1b. The final trailing spoiler gradients are also of the same design, although its polarity is adjusted to maintain alternating polarity within that axis.

Fig. 1
figure 1

a. 6 pulse train numerically optimized for 3 T (\( {\widehat{B}}_1 \) = 0.4–1.2, B0 = ±240 Hz, B1 field = 14 μT), with a total duration of 32.8 ms. Gradient areas are the effective areas across all three gradient directions. b. 4 pulse train numerically optimized for 1.5 T (\( {\widehat{B}}_1 \) = 0.7–1.0, B0 = ±120 Hz, B1 field = 26.9 μT), with a total duration of 23.6 ms. The first four spoilers are the same for the 4 and 6 pulse trains and the last spoilers have the same pattern, but with the polarity in the slice select axis reversed to maintain alternating polarity. c. The optimized BIR4-90 pulse, with a total duration of 12.2 ms. All saturation pulses are shown on the same time, RF amplitude and gradient strength scale

Numerical pulse train design

The residual MZ/M0 following pulse train saturation is dependent on the flip angle order due to T1 effects and thus direct gradient descent minimization of an ordered flip angle set may be trapped in local minima. Instead, an unordered set of flip angles was minimized using a standard Nelder-Mead algorithm [12], but in each iteration, all possible permutations of the flip angle set were evaluated and the lowest |MZ/M0| from all permutations was returned. A detailed description of the minimization algorithm can be found in Additional file 1 and MATLAB source code used is provided in Additional files 2, 3, 4, 5 and 6 and available online at https://bitbucket.org/kelvinc/pulsetrainopt.

Optimal flip angles were calculated for pulse trains ranging from 3 to 6 RF pulses. Pulse trains were optimized for 1.5 T using a \( {\widehat{B}}_1 \) range of 0.7–1.0 [13], a B0 range corresponding to ±120 Hz, and a maximum B1 field strength of 26.9 μT. Pulse trains were also optimized for 3 T using a larger \( {\widehat{B}}_1 \) range of 0.4–1.2 [8, 14], a B0 range corresponding to ±240 Hz, and a maximum B1 of 14 μT. In all Bloch simulation optimizations, the T1 range was 200–2000 ms and T2 was 45 ms (native myocardium). T2 dependence of optimized pulse trains was characterized through simulations using T2 values from 5 ms to 250 ms.

Proposed saturation pulse trains were compared to a commonly used reference pulse train consisting of three 90° RF pulses, each with a fixed duration of 0.5 ms [7]. Spoiler durations for the reference pulse train were 1.00, 8.80, 5.86, and 1.55 ms, with a total time of 18.71 ms.

BIR4-90 pulse optimization

A BIR4-90 pulse was numerically optimized by a brute force optimization over adiabatic design parameters [15]. The pulse duration, amplitude, maximum frequency sweep, and parameters ζ and κ which define the adiabatic half passage section of the BIR4-90 defined in Equations 4 and 6 in [15] were varied and the optimum pulse was selected which achieved the best saturation over the specified range of \( {\widehat{B}}_1 \) and off-resonance, i.e., best worst-case deviation. The frequency sweep parameter was varied from 4–20 kHz in 1 kHz steps, tan(κ) ranged from 8 to 22 in steps of 2, and ζ ranged from 8 to 30 in steps of 2. Bloch simulations assumed values for native myocardium of T1 = 1100 ms and T2 = 45 ms. Pulse durations between 4 and 10 ms were evaluated. The design range was ±120 Hz off-resonance and \( {\widehat{B}}_1 \) ranged from 0.7 to 1.0. The peak amplitude was limited to approximately 20 μT. The optimized BIR4-90 pulse was further characterized with simulations using T2 values from 5 ms to 250 ms.

RF power calculation

Relative RF energy was calculated for each saturation pulse as B1 2 integrated over time divided by the RF energy of the 90°-90°-90° pulse train. Relative peak RF power was calculated as the maximum B1 2 for each saturation pulse, normalized to the maximum B1 2 in the reference 90°-90°-90° pulse train.

Phantom validation

Saturation efficiency was experimentally assessed for five saturation pulses: the reference 90°-90°-90° pulse train, the 3 T optimized 6-pulse train (Fig. 1a), the 1.5 T optimized 4-pulse train (Fig. 1b), the reference BIR4-90 pulse, and the optimized BIR4-90 pulse (Fig. 1c). Performance for each pulse was assessed in a Siemens 2 L plastic bottle phantom (1.25 g NiSO46H2O + 5 g NaCl per 1000 g water). Imaging was performed on a Siemens 1.5 T MAGNETOM Aera MRI scanner (Siemens Healthcare; Erlangen, Germany) and the pulses in the 3 T 6-pulse train were limited to 14 μT to reflect the typical limitation at 3 T. The T1 value was measured using an inversion recovery spin echo sequence with inversion times between 100 ms and 500 ms and a non-prepared image. T2 was measured using spin echo repeated with echo times varied between 11 ms and 300 ms. All spin echo imaging was performed with a single k-space line acquired per excitation and a 10 s TR to ensure complete magnetization recovery between excitations.

A linear spatial gradient was applied along the long axis of the phantom to assess performance as a function of off-resonance. To calculate an off-resonance (B0) map, gradient echo images were acquired with 5 echo times (TE) between 4 and 6 ms, 3000 ms repetition time (TR), 90° flip angle, 128 × 44 matrix size, 400 × 138 mm2 field of view, and 8 mm slice thickness. The B0 map was generated by fitting the equation φ = ΔB0TE for every pixel, where φ is the phase of gradient echo images. To calculate an excitation \( {\widehat{B}}_1 \)-field map for the phantom, gradient echo images were acquired with 45°, 60°, 90°, and 120° flip angles, 3.79 ms TE, and other parameters as above. The signal intensity as a function of flip angle was simulated using a Bloch equation simulation accounting for slice profile effects [16], and a \( {\widehat{B}}_1 \) map was generated by fitting the measured signal intensity to this simulated curve for every pixel.

Saturation pulse efficiency was calculated using a gradient echo sequence with 3.79 ms TE, 3000 ms TR, 90° flip angle, and other parameters as above. Images were acquired without a saturation pulse and with 4 saturation recovery times (TS = 10–40 ms) for each of the 5 saturation pulses, where TS is defined as the time from the end of the last RF pulse in the saturation and the middle of the imaging excitation RF pulse. A saturation efficiency map was generated by fitting acquired data to {Signal = A*[1 - ηexp(−TS/T1)]}, where η is the saturation efficiency and residual (signed) MZ/M0 is equal to 1-η. These acquisitions were repeated 9 times with the flip angles of the saturation RF pulses scaled by 40–120 % to emulate the effect of varied \( {\widehat{B}}_1 \) values.

For each pixel of the nine resulting residual MZ/M0 maps, the B0 and effective \( {\widehat{B}}_1 \) (equal to \( {\widehat{B}}_1 \) map multiplied by the flip angle scaling factor) could be determined, separately for each saturation pulse considered. Together, the intrinsic variation in \( {\widehat{B}}_1 \) over the phantom in combination with the 9 repeated experiments with flip angle scaling from 40 % to 120 % yielded a wide range of effective \( {\widehat{B}}_1 \) values. With the variation in B0 provided by the linear gradient, the MZ/M0 values were thus directly measured over a wide range of B0 and \( {\widehat{B}}_1 \) values. The missing data within the B0\( {\widehat{B}}_1 \) space was calculated by fitting a surface over the scattered points with a modified ridge estimator using the gridfit function [17].

SASHA sequence simulation

Bloch equation simulations of the SASHA sequence were performed with a range of residual MZ/M0 (−5 % to 5 %) and relaxation values emulating myocardium (native T1/T2: 1175/50 ms, post-contrast T1/T2: 725/50 ms) and blood (native T1/T2: 1650/240, post-contrast: 500/180 ms) at 1.5 T [3]. Typical in-vivo SASHA acquisition parameters were used: 1.3/2.6 ms TE/TR, 70° target flip angle, 70 phase encode lines, 60 bpm simulated heart rate, 10 images with maximum 775 ms saturation recovery time (TS) and a non-saturated image. Image readout flip angles were scaled using a variable flip angle (VFA) scheme to minimize T1 error with two parameter fitting as previously described [18]. In this scheme, the prescribed flip angle of the first 45 RF pulses was scaled by sin(x) for π/90 < x < π/2. Simulations used actual RF pulse waveforms and the transverse magnetization (MXY) at the centre k-space readout as the image signal intensity. T1 values were calculated by fitting simulated data to a 2-parameter exponential recovery model with assumed ideal saturation.

In vivo human study

Saturation efficiency was assessed in 3 human volunteers on a Siemens 3 T MAGNETOM Skyra MRI scanner (Siemens Healthcare; Erlangen, Germany) using a single-shot centrically ordered gradient echo sequence. Volunteers provided written informed consent and the study was approved by the University of Virginia Health System Institutional Review Board. Typical sequence parameters were: 1.14/2.31 ms TE/TR, 10° flip angle, 128 × 90 matrix, 400 × 281 mm2 field of view, and 10 mm slice thickness. Saturation recovery images were acquired for the 90°-90°-90°, 1.5 T 4-pulse, 3 T 6-pulse, reference BIR4-90, and optimized BIR4-90 saturation pulses with a 7.5 ms saturation recovery time, defined from the end of the last saturation RF pulse to the center of the first imaging RF pulse. A proton density image was acquired with the same pulse sequence parameters. All images were obtained in short-axis and 4-chamber orientations, acquired in diastasis with end-expiratory breath-holds.

Saturation recovery images were processed using a phase-sensitive reconstruction using the proton density image phase as a reference in order to avoid Rician noise bias. The residual MZ/M0 was calculated as the ratio of the saturation recovery image to the proton density image.

\( {\widehat{B}}_1 \) maps were acquired using a free-breathing saturated double angle method [11] with single-shot echo planar imaging (EPI) readouts. Typical sequence parameters were: electrocardiogram (ECG) gated diastasis imaging, 45°/90° flip angles, 80 × 50 matrix, rate 2 parallel imaging (GRAPPA), 360 × 225 mm2 field of view, 10 mm slice thickness, 2 saturation bands on the chest and back to reduce the field of view, and 15 repetitions per flip angle. All images were registered and a \( {\widehat{B}}_1 \) scale factor map was calculated as \( {\widehat{B}}_1={ \cos}^{-1}\left(\left|\frac{I_{90{}^{\circ}}}{2{I}_{45{}^{\circ}}}\right|\right)/45{}^{\circ} \), where I45° and I90° are the signal intensities from the 45° and 90° flip angle images respectively.

T1 maps were collected using SASHA with 90°-90°-90°, optimized BIR4-90, and 6-pulse saturation and MOLLI with an optimized adiabatic inversion pulse [13] in diastasis with end-expiratory breath-holds. SASHA sequences used a 70° imaging flip angle with VFA modulation and 10 images with the longest TS time allowable by the heart rate [19] and MOLLI used a 35° imaging flip angle with a 5(3)3 sampling pattern.

In vivo swine study

Myocardial T1 mapping was performed on a 66 kg swine on a Siemens 3 T Skyra MRI scanner using the SASHA and MOLLI [20] T1 mapping sequences. SASHA was acquired using both 90°-90°-90° and 6-pulse saturation with a 45° imaging flip angle and VFA modulation. MOLLI used an optimized inversion pulse with a 20° imaging flip angle. A \( {\widehat{B}}_1 \) map was acquired using a saturated double angle method as described above. Swine were studied in accordance with a protocol approved by the Animal Care and Use Committee of the National Institutes of Health (protocol number H-0214).

Results

B1 field strength limitations

The maximum B1 field strength was calculated for 379 subjects on the Siemens MAGNETOM Skyra platform and 230 subjects on the Siemens MAGNETOM Aera platform. A maximum B1 field strength of 14 μT was achievable on the 3 T Skyra platform in >98 % of subjects, while 26.9 μT was achievable on the 1.5 T Aera platform in >99 % of subjects. At a B1 strength of 14 μT, a 90° rectangular pulse is 419 μs and the full width half maximum effective B1 scaling is ±1334 Hz.

BIR4-90 pulses

The numerically optimized BIR4-90 design used a duration of 5.12 ms, maximum B1 field of 20.6 μT, swept over ±7 kHz, with ζ = 22 and tan(κ) = 18. This design is longer and has a lower maximum B1 compared to the reference BIR4-90 pulse that has a 4.00 ms duration and a maximum B1 of 25.1 μT, resulting in lower relative peak RF energy and power (Table 1). The optimized BIR4-90 design had substantially improved performance with a maximum residual |MZ/M0| within the 1.5 T optimization range 5.8× lower than the reference pulse. More detailed comparisons of the two BIR4-90 pulses are characterized in the phantom study below.

Table 1 Summary of characteristics for Bloch simulation numerically optimized saturation pulses

Numerically optimized pulse trains

Characteristics of the Bloch simulation numerically optimized pulse trains are summarized in Table 1 and their simulated performance is plotted in Fig. 2. The reference 90°-90°-90 pulse train has poor performance at the lower end of \( {\widehat{B}}_1 \) scale factors expected at 1.5 T, with a residual MZ/M0 of 10 % at a \( {\widehat{B}}_1 \) of 0.7 while on-resonance and a 1000 ms T1. For the 1.5 T optimization range, the optimized 3-pulse train (90°-107°-126°) achieved the target residual |MZ/M0| of <1 %, while 6 pulses (115°-90°-125°-85°-176°-223°) were necessary to achieve this target for the 3 T optimization range due to the larger \( {\widehat{B}}_1 \) range of 0.4–1.2. For both \( {\widehat{B}}_1 \) optimization ranges, simulated residual |MZ/M0| increased rapidly outside of the optimization region. Minor off-resonance and T1 effects were also observed, but did not have a consistent trend over the range (Fig. 2).

Fig. 2
figure 2

Simulated performance for pulse trains with 3–6 pulses, visualized for 3 T1 values (colored lines) and off-resonance (B0) values (solid vs. dashed lines). Performance for the reference 90°-90°-90° pulse train (on-resonance, 1000 ms T1) is also shown along with the 3 pulse train optimized for 1.5 T for comparison (top left)

Pulse train ordering effects

The saturation efficiency of a saturation pulse train is dependent on the order of its flip angles due to T1 relaxation. The effect of flip angle order in pulse trains can be illustrated by characterizing the residual |MZ/M0| for all permutations of a set of flip angles. For the 6-pulse train optimized for 3 T (Table 1), the residual |MZ/M0| was calculated for all 720 flip angle permutations during the final iteration of the Bloch simulation numerical optimization algorithm. The maximum and mean residual |MZ/M0| are calculated over the optimization range of B0, \( {\widehat{B}}_1 \), and T1 values and plotted (Additional file 7) after sorting by the maximum residual |MZ/M0|. There was relatively poor correlation between the maximum and mean |MZ/M0|, as some permutations with similar maximum |MZ/M0| had greatly different mean |MZ/M0|. The largest maximum residual |MZ/M0| was 5.7× greater than the smallest, and the largest mean residual |MZ/M0| was 2.6× greater than the smallest.

T1 relaxation effects

Simulated residual MZ/M0 for selected saturation pulses are shown as a function of \( {\widehat{B}}_1 \) and T1 in Fig. 3. A larger range of \( {\widehat{B}}_1 \) values (0.4 and 1.2) are shown only for the 6-pulse train, which is full range for the 3 T field strength. Performance is only weakly related to T1 in the optimization range used for pulse design (200–2000 ms), but residual |MZ/M0| increases rapidly for T1 values shorter than 25 ms for all pulse trains, with divergent behavior depending on the \( {\widehat{B}}_1 \) scale factor. The 90°-90°-90° train has poor performance with a \( {\widehat{B}}_1 \) scale factor of 0.7, with residual |MZ/M0| exceeding 15 % for a 50 ms T1. The 3 T 6-pulse train maintains |MZ/M0| < 1 % for T1 greater than 186 ms at ±240 Hz and for T1 greater than 140 ms at ±120 Hz. The optimized BIR4-90 pulse had the best performance at short T1 values, with 1.6 % residual |MZ/M0| at 25 ms.

Fig. 3
figure 3

Simulated residual MZ/M0 for selected saturation pulses as a function of T1 for different off-resonance (B0) and \( {\widehat{B}}_1 \) values

T2 relaxation effects

The effects of T2 relaxation on residual MZ/M0 are shown for the optimized 3 T 6-pulse train and BIR4-90 saturation pulses in Fig. 4. Saturation pulse trains were insensitive to T2 relaxation and the optimized BIR4-90 pulse had less than a 1 % change in residual MZ/M0 for T2 values from 25 to 250 ms.

Fig. 4
figure 4

Simulated residual MZ/M0 for selected saturation pulses as a function of T2 for different T1, off-resonance (B0) and \( {\widehat{B}}_1 \) values

Phantom validation

The T1 and T2 of the bottle phantom was measured to be 316 ms and 274 ms respectively using spin echo experiments, and the intrinsic \( {\widehat{B}}_1 \) scaling factor across the image was found to range from 0.83 to 1.05. The applied linear magnetic field gradient resulted in off-resonance values from −605 to 697 Hz across the phantom. The performance of each saturation pulse as a function of B0 and \( {\widehat{B}}_1 \) scaling factor is plotted in the top row of Fig. 5, with the optimization range of B0 and \( {\widehat{B}}_1 \) values demarcated with a white box.

Fig. 5
figure 5

Simulated residual MZ/M0 (top row) and experimentally measured values (middle row) for 5 different saturation pulses as a function of \( {\widehat{B}}_1 \) scale factor and off-resonance. The bottom row shows a profile of measured residual MZ/M0 as a function of \( {\widehat{B}}_1 \) scale factor for three off-resonance frequencies

Experimentally measured residual MZ/M0 (middle row, Fig. 5) shows excellent agreement with simulations (top row, Fig. 5). Saturation pulse trains are robust to off-resonance due to the short individual pulse durations, while the BIR4-90 pulses with an overall shorter duration have tighter off-resonance constraints and more complex relationships with B0 and \( {\widehat{B}}_1 \). The reference 90°-90°-90 pulse train shows poor performance at low \( {\widehat{B}}_1 \) scaling factors, even in the relatively narrow range of 0.7–1.0 expected at 1.5 T. All proposed pulses show greatly improved performance within their respective optimization spaces (white dashed box) compared to the reference designs. The average measured residual |MZ/M0| within these spaces was 0.24 ± 0.16 %, 0.11 ± 0.06 %, and 0.20 ± 0.18 % for the optimized BIR4-90, 4 pulse (1.5 T), and 6 pulse (3 T) trains respectively.

SASHA sequence simulation

Figure 6 shows the relationship between residual MZ/M0 and simulated T1 errors in the SASHA sequence. SASHA T1s are generally underestimated with incomplete saturation, with a 1.3–1.6 % increase in 2-parameter SASHA T1 error for every 1 % change in residual MZ/M0. A higher heart rate of 100 bpm results in greater SASHA T1 errors for native (pre-contrast) T1/T2 values compared to 60 bpm, with negligible effects for post-contrast values. Overall, 2-parameter model fitting results in greater T1 error dependence on residual MZ/M0 than 3-parameter fitting, particularly for the lower post-contrast T1/T2 values.

Fig. 6
figure 6

Simulated T1 error for SASHA as a function of residual MZ/M0 for native (top) and post-contrast (bottom) myocardium (left) and blood (right) using 2 and 3 parameter models at 60 and 120 bpm. Input native relaxation times were T1/T2 = 1175/50 ms (myocardium) and T1/T2 = 1650/240 (blood) and post-contrast relaxation times were T1/T2 = 725/50 ms (myocardium) and T1/T2 = 500/180 ms (blood)

In vivo human study

Subjects had an average heart rate of 63 ± 10 bpm with no mis-triggering. Representative \( {\widehat{B}}_1 \) and residual MZ/M0 maps from one subject are shown in Fig. 7. \( {\widehat{B}}_1 \) was decreased near the acute margin of the right ventricle and coincided with increased residual MZ/M0 values for the 90°-90°-90° and reference BIR4-90 pulses. A similar spatial pattern was observed in all subjects, although there was considerable variability in \( {\widehat{B}}_1 \) between subjects (Table 2). The optimized BIR4-90 pulse had lower residual MZ/M0 than the reference BIR4-90 pulse in the myocardium (2.8 ± 1.1 % vs 6.1 ± 3.1 %) as well as in the blood pool. The 6-pulse train had the lowest overall residual MZ/M0 in the myocardium (1.8 ± 0.3 %) and significantly more uniform saturation over the entire body (Fig. 7).

Fig. 7
figure 7

\( {\widehat{B}}_1 \) scale factor and residual MZ/M0 maps measured in a human volunteer for various saturation pulses. Air regions with low signal intensity are masked for visualization. A profile is extracted from the short axis images (black and white dashed line) across both the right and left ventricles

Table 2 Measured residual longitudinal magnetization (MZ/M0) at 3 T in 3 human subjects

Calculated in-vivo T1 values for the SASHA sequence with various saturation pulses and MOLLI are summarized in Table 3. The 6-pulse SASHA sequence had longer myocardial and blood T1 values than the 90°-90°-90° and optimized BIR-4 variants and all SASHA T1 values were longer than MOLLI T1 values. Right ventricular blood T1s were lower than left ventricular blood T1s for all sequences, with the largest differences in the 90°-90°-90° SASHA and MOLLI data (193 ± 100 ms and 143 ± 83 ms respectively).

Table 3 Measured T1 values at 3 T in 3 human subjects

In vivo swine study

Parametric T1 and \( {\widehat{B}}_1 \) maps acquired in a swine are shown in Fig. 8 and a profile is extracted along the left ventricle. Myocardial T1 values from the 90°-90°-90° SASHA and MOLLI sequences show a >50 % artifactual decrease in the lateral wall, spatially coinciding with reduced \( {\widehat{B}}_1 \) values. T1 values using the 6-pulse SASHA sequence were more spatially uniform across both the myocardium (1386 ± 70 ms along the profile) and the overall field of view.

Fig. 8
figure 8

A \( {\widehat{B}}_1 \) map (bottom left) and T1 maps (top row) using SASHA with a reference 90°-90°-90° saturation pulse, SASHA with a proposed 6-pulse train, and MOLLI in a swine. Air regions with low signal intensities are masked for visualization. A profile of T1 values along the left ventricular wall shows decreased T1 values in the lateral wall coinciding with reduced \( {\widehat{B}}_1 \) values

Discussion

In this study, adiabatic BIR4-90 and pulse train saturation designs were optimized for 1.5 T and 3 T with the objective of achieving <1 % residual |MZ/M0| for the wide ranges of \( {\widehat{B}}_1 \) and B0 inhomogeneities expected at these field strengths, while imposing the typical peak B1 field amplitude limitations measured on 1.5 T and 3 T scanners. New pulses were shown to have significantly improved saturation performance as compared to current standard pulses. Measured saturation performance in phantoms had excellent agreement with simulations and in-vivo experiments demonstrated the superior performance of the proposed 6-pulse train saturation pulse. T1 maps calculated using SASHA with proposed 6-pulse train saturation showed greater spatial homogeneity than the reference pulse in both human and swine studies.

BIR4-90 pulse design

Experimental data in this study shows the reference BIR4-90 saturation pulse has relatively poor performance with low \( {\widehat{B}}_1 \) and high off-resonance within the range expected at 1.5 T. A new optimized BIR4-90 pulse was shown to have improved performance with <1 % residual |MZ/M0| over the 1.5 T optimization range of B0 and \( {\widehat{B}}_1 \) fields while reducing maximum B1 field strength and resulting SAR. Although the required B1 of 20.6 μT was achievable on the 1.5 T MRI systems used here, the 14 μT B1 limit at 3 T reduces the effective flip angle and saturation performance. In the 3 T human study, the optimized BIR4-90 SASHA had 1.6 % shorter myocardial T1 values than 6-pulse SASHA, consistent with T1 underestimation from slightly larger residual MZ/M0.

Pulse train design

The ordering of flip angles in saturation pulse trains was found to be important, with a reduction in the maximum residual |MZ/M0| by a factor of 5.7× from the best order permutation to the worst. The evaluation of all permutations of an unordered flip angle set in the minimization subroutine provided a straightforward means of considering flip angle ordering while avoiding trapping within local minima. This is likely to provide a more optimal solution than the previous minimization approach which considered ordering only in the final stage [8].

Pulse trains were designed for expected ranges of \( {\widehat{B}}_1 \) values at both 1.5 T and 3 T, with excellent performance within those optimization ranges when a sufficient number of pulses are used. As performance decreases rapidly outside of the \( {\widehat{B}}_1 \) optimization range for all pulse trains (Fig. 2), conservative estimates of \( {\widehat{B}}_1 \) ranges were used in the design process. Experimentally measured residual MZ/M0 in a phantom closely matched simulations for both the adiabatic BIR4-90 pulses as well as all saturation pulse trains, confirming the validity of the Bloch equation simulations used in the design process.

For the \( {\widehat{B}}_1 \) range of 0.7 to 1.0 expected for 1.5 T, the optimized 3-pulse train has the same number of pulses as the reference 90°-90°-90° pulse train and reduces the maximum residual |MZ/M0| by 14×, meeting the target of <1 %. The 4-pulse train has even better performance with <0.25 % peak residual magnetization (0.11 % mean residual signal), with minimal additional time and SAR.

The design \( {\widehat{B}}_1 \) range for 3 T was chosen to be 0.4–1.2 to encompass the widest range of in-vivo cardiac \( {\widehat{B}}_1 \) values measured in a handful of small studies [8, 11, 14, 21]. The in-vivo \( {\widehat{B}}_1 \) maps for the 3 human subjects in this study show a smaller range of \( {\widehat{B}}_1 \) values than previous studies, but with a similar spatial pattern of variation primarily in the anteroposterior direction and larger \( {\widehat{B}}_1 \) values in the lateral wall of the left ventricle [14]. This may be due to a difference in body coil design between scanner manufacturers, and further variations may exist between models with different bore sizes. However, \( {\widehat{B}}_1 \) values less than 0.4 were observed in the swine model, suggesting that the design \( {\widehat{B}}_1 \) range may need to be further widened for large animal model studies. Further work is needed to characterize the ranges of \( {\widehat{B}}_1 \) values in larger representative populations of both human and swine on different scanner platforms.

The 3 T 6-pulse train achieved <1 % peak residual magnetization over the wide design range of \( {\widehat{B}}_1 \) values, with excellent performance (mean residual |MZ/M0| of 0.27 %) and only 1.4× greater peak RF power compared to the 90°-90°-90° reference pulse train. The relatively short RF pulse durations used in the pulse trains resulted in a wide range of B0 insensitivity for all pulse train designs, with a negligible drop in measured performance at ±450 Hz (Fig. 5). This is consistent with superior saturation of fat signals, which have a ~440 Hz shift in resonant frequency at 3 T, and more uniform saturation over the entire field of view compared to the BIR4-90 pulses (Fig. 7).

Further pulse train optimization

The saturation pulses presented in this study were optimized for myocardial T1 mapping, but their excellent performance over a typical optimization range makes them suitable for other applications as well such as myocardial first-pass perfusion sequences or the saturated double angle method. In settings with much different criteria, such as B1 amplitude limitations or B0, \( {\widehat{B}}_1 \), and T1 ranges, the supplied optimization code may be used to design more application-specific pulse trains.

For example, while relatively short RF pulses were selected to maximize the off-resonance performance, pulse lengths may be increased considerably to reduce peak B1 amplitude and SAR. Additionally, while the maximum B1 amplitude was characterized for 2 common MRI scanner models, other models and scanners by different vendors likely have different limitations. The inter-pulse spoiler durations could also be shortened to further reduce T1 dependencies, which would improve performance for very short T1 values encountered in contrast-enhanced first-pass myocardial perfusion imaging.

Saturation performance and SASHA T1 mapping

Simulations of 2-parameter fitting with SASHA data show increased sensitivity to saturation pulse performance compared to 3-parameter fitting. In human studies, there was consistently lower \( {\widehat{B}}_1 \) in the right ventricle leading to increased residual longitudinal magnetization with the 90°-90°-90° saturation pulse, but not the 6-pulse train. Correspondingly, right ventricular T1 values with 90°-90°-90° SASHA were substantially lower (193 ± 100 ms) than in the left ventricle, but the difference was much lower (57 ± 13 ms) with 6-pulse SASHA which had more spatially uniform saturation. This remaining difference between the left and right ventricular blood with 6-pulse SASHA is comparable to a difference of ~80 ms between atrial and venous blood with a 0.42 hematocrit previously reported at 3 T [22]. Shorter MOLLI T1 values compared to SASHA are consistent with known influences of factors such as T2, magnetization transfer, and off-resonance on the MOLLI sequence [4].

Substantially larger \( {\widehat{B}}_1 \) variations observed in the swine may be the result of the more cylindrical chest geometry compared to humans. These variations resulted in a >50 % artifactual decrease in myocardial T1 values for both the 90°-90°-90° SASHA and MOLLI sequences while 6-pulse SASHA had spatially homogenous T1s. SASHA with a 6-pulse saturation may provide more reliable T1 measurements in pre-clinical large animal studies at 3 T, but further study is needed to determine the impact on the quantification of myocardial fibrosis.

Study limitations

Gradient spoiler design and its effect on the possible formation of stimulated echoes were not studied in detail. The proposed spoiler design utilizes relatively large gradient areas played in multiple directions, with care taken to avoid duplicate gradient areas. While no stimulated echo artifacts were observed in our data, a more systematic study of stimulated echo formation is warranted.

Residual MZ/M0 was measured in-vivo using images with a saturation recovery time of 7.5 ms, which was the minimal duration due to the length of post-saturation spoiler gradient for the reference BIR4-90 pulse. Calculated residual MZ/M0 values are likely to have a small positive bias due to T1 recovery during this duration. Although centric ordering was used, T1 recovery during the readout results in a further positive bias and may explain the larger residual MZ/M0 in the LV myocardium compared to the blood pool it encloses.

In-vivo measurements in this study were limited to 3 human subjects and a single animal. However, the magnitude and spatial pattern of \( {\widehat{B}}_1 \) heterogeneity in the human subjects are consistent with existing literature [14] and \( {\widehat{B}}_1 \) variations in the swine were representative of the authors’ previous experience.

Conclusions

This study has described the optimization of saturation pulses and validated their performance in both phantom experiments and in-vivo studies. While an optimized BIR4-90 design and a 3-pulse train were able to meet the design criteria of <1 % residual |MZ/M0| at 1.5 T, only the optimized 6-pulse train met the criteria for B0 and \( {\widehat{B}}_1 \) values expected at 3 T. The robust performance of optimized saturation pulse trains to a wide range of off-resonance, T1 values, and \( {\widehat{B}}_1 \) scale factors can minimize these as sources of errors in quantitative saturation-recovery based sequences such as SASHA or in other applications such as first pass perfusion imaging.

Abbreviations

ECG:

Electrocardiogram

GRAPPA:

Generalized auto-calibrating partially parallel acquisition

LV:

Left ventricle

MOLLI:

Modified Look-Locker inversion recovery

RF:

Radiofrequency

RV:

Right ventricle

SAR:

Specific absorption rate

SASHA:

Saturation recovery single-shot acquisition

TE:

Echo time

TR:

Repetition time

TS:

Saturation recovery time

VFA:

Variable flip angle

References

  1. Wacker CM, Bock M, Hartlep AW, Beck G, van Kaick G, Ertl G, et al. Changes in myocardial oxygenation and perfusion under pharmacological stress with dipyridamole: assessment using T*2 and T1 measurements. Magn Reson Med. 1999;41:686–95.

    Article  CAS  PubMed  Google Scholar 

  2. Higgins DM, Ridgway JP, Radjenovic A, Sivananthan UM, Smith MA. T1 measurement using a short acquisition period for quantitative cardiac applications. Med Phys. 2005;32:1738–46.

    Article  PubMed  Google Scholar 

  3. Chow K, Flewitt JA, Green JD, Pagano JJ, Friedrich MG, Thompson RB. Saturation recovery single-shot acquisition (SASHA) for myocardial T1 mapping. Magn Reson Med. 2014;71:2082–95.

    Article  PubMed  Google Scholar 

  4. Kellman P, Hansen MS. T1-mapping in the heart: accuracy and precision. J Cardiovasc Magn Reson. 2014;16:2.

    Article  PubMed Central  PubMed  Google Scholar 

  5. Ogg RJ, Kingsley PB, Taylor JS. WET, a T1- and B1-insensitive water-suppression method for in vivo localized 1H NMR spectroscopy. J Magn Reson B. 1994;104:1–10.

    Article  CAS  PubMed  Google Scholar 

  6. Tao Y, Hess AT, Keith GA, Rodgers CT, Liu A, Francis JM, et al. Optimized saturation pulse train for human first-pass myocardial perfusion imaging at 7 T. Magn Reson Med 2015;73:1450–56.

    Article  PubMed Central  PubMed  Google Scholar 

  7. Oesingmann N, Zhang Q, Simonetti O. Improved saturation RF pulse design for myocardial first-pass perfusion at 3 T. J Cardiovasc Magn Reson. 2004;6:373–4.

    Google Scholar 

  8. Sung K, Nayak KS. Design and use of tailored hard-pulse trains for uniformed saturation of myocardium at 3 Tesla. Magn Reson Med. 2008;60:997–1002.

    Article  PubMed  Google Scholar 

  9. Staewen RS, Johnson AJ, Ross BD, Parrish TB, Merkle H, Garwood M. 3-D FLASH imaging using a single surface coil and a new adiabatic pulse, BIR-4. Invest Radiol. 1990;25:559–67.

    Article  CAS  PubMed  Google Scholar 

  10. Kim D, Oesingmann N, McGorty K. Hybrid adiabatic-rectangular pulse train for effective saturation of magnetization within the whole heart at 3 T. Magn Reson Med. 2009;62:1368–78.

    Article  PubMed Central  PubMed  Google Scholar 

  11. Cunningham CH, Pauly JM, Nayak KS. Saturated double-angle method for rapid B1+ mapping. Magn Reson Med. 2006;55:1326–33.

    Article  PubMed  Google Scholar 

  12. Nelder JA, Mead R. A Simplex Method for Function Minimization. Comput J. 1965;7:308–13.

    Article  Google Scholar 

  13. Kellman P, Herzka DA, Hansen MS. Adiabatic inversion pulses for myocardial T1 mapping. Magn Reson Med. 2014;71:1428–34.

    Article  PubMed Central  PubMed  Google Scholar 

  14. Sung K, Nayak KS. Measurement and characterization of RF nonuniformity over the heart at 3 T using body coil transmission. J Magn Reson Imaging. 2008;27:643–8.

    Article  PubMed  Google Scholar 

  15. Garwood M, Ke Y. Symmetric pulses to induce arbitrary flip angles with compensation for rf inhomogeneity and resonance offsets. J Magn Reson 1991;94:511–25.

    Google Scholar 

  16. Wang J, Mao W, Qiu M, Smith MB, Constable RT. Factors influencing flip angle mapping in MRI: RF pulse shape, slice-select gradients, off-resonance excitation, and B0 inhomogeneities. Magn Reson Med. 2006;56:463–8.

    Article  PubMed  Google Scholar 

  17. Surface Fitting using gridfit [http://www.mathworks.com/matlabcentral/fileexchange/8998-surface-fitting-using-gridfit]. Accessed June 14, 2014.

  18. Chow K, Spottiswoode B, Pagano JJ, Thompson R. Improved precision in SASHA T1 mapping with a variable flip angle readout. J Cardiovasc Magn Reson. 2014;16:M9.

    Article  PubMed Central  Google Scholar 

  19. Kellman P, Xue H, Chow K, Spottiswoode BS, Arai AE, Thompson RB. Optimized saturation recovery protocols for T1-mapping in the heart: influence of sampling strategies on precision. J Cardiovasc Magn Reson. 2014;16:55.

    Article  PubMed Central  PubMed  Google Scholar 

  20. Messroghli DR, Radjenovic A, Kozerke S, Higgins DM, Sivananthan MU, Ridgway JP. Modified Look-Locker inversion recovery (MOLLI) for high-resolution T1 mapping of the heart. Magn Reson Med. 2004;52:141–6.

    Article  PubMed  Google Scholar 

  21. Rao AK, Greve AM, Nielles-Vallespin S, Spottiswoode BS, Chow K, Thompson RB, et al. Variability of T1 in purpose recruited normal volunteers and patients as a function of shim (B0), flip angle (B1) and myocardial sector at 3 T. J Cardiovasc Magn Reson. 2015;17:P5.

    Article  PubMed Central  Google Scholar 

  22. Lu H, Clingman C, Golay X, van Zijl PCM. Determining the longitudinal relaxation time (T1) of blood at 3.0 Tesla. Magn Reson Med. 2004;52:679–82.

    Article  PubMed  Google Scholar 

Download references

Funding

This work was supported by the Canadian Institutes of Health Research, Alberta Innovates – Health Solutions, and National Heart, Lung and Blood Institute, National Institutes of Health by the Division of Intramural Research, NHLBI, NIH, DHHS (HL004607-14CPB).

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Richard B. Thompson.

Additional information

Competing interests

Dr. Spottiswoode is employed by Siemens Medical Solutions USA, Inc. Dr. Arai is a principal investigator on a US government Cooperative Research And Development Agreement (CRADA) with Siemens Medical Solutions (HL-CR-05-004).

Authors’ contributions

KC, PK, and RBT contributed to study design and saturation pulse optimization and BSS assisted in pulse sequence implementation. KC performed pulse train simulations, phantom experiments and all data analysis. SNV and AEA were responsible for swine imaging and the \( {\widehat{B}}_1 \) mapping sequence and MS assisted in human imaging. All authors contributed to and approved the final manuscript.

Additional files

Additional file 1:

Numerical pulse train optimization algorithm. (DOCX 137 kb)

Additional file 2:

Main optimization code (MATLAB). (TXT 5 kb)

Additional file 3:

Additional optimization code (MATLAB). (TXT 2 kb)

Additional file 4:

Additional optimization code (MATLAB). (TXT 2 kb)

Additional file 5:

Additional optimization code (MATLAB). (TXT 2 kb)

Additional file 6:

Additional optimization code (MATLAB). (TIFF 4449 kb)

Additional file 7:

Saturation Pulse Performance with Flip Angle Permutations (3 T 6-pulse train). (TXT 8 kb)

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Chow, K., Kellman, P., Spottiswoode, B.S. et al. Saturation pulse design for quantitative myocardial T1 mapping. J Cardiovasc Magn Reson 17, 84 (2015). https://doi.org/10.1186/s12968-015-0187-0

Download citation