Phantom validation of 4D flow: independent validation of flow velocity quantification using particle imaging velocimetry
Journal of Cardiovascular Magnetic Resonance volume 17, Article number: O18 (2015)
Background
4D phase contrast magnetic resonance blood flow measurements (4D PC-MR) is a promising tool for clinical assessment of cardiac function, but lack of independent validation may limit clinical adoption. Previous validation studies are limited by lack of pulsatility [1, 2] or by the need of a post hoc correction factor to compensate for differences in experimental setup between 4D PC-MR and reference imaging [3]. Therefore, we aimed to validate 4D PC-MR velocity measurements with particle imaging velocimetry (PIV) as the reference standard in a phantom setup based on vortex ring formation, which has previously shown excellent repeatability [4].
Methods
We constructed a pulsatile pump and a water tank with a 25mm nozzle (Figure 1). Five different pump settings were used. PIV measurements with spatial resolution 1.5x1.5 mm were obtained on two consecutive days to evaluate phantom stability. Nozzle stroke volumes were measured using 2D PC-MR (voxel size 2.4x2.4x6 mm, VENC 50 cm/s). 4D PC-MR (3x3x3 mm voxels, temporal resolution 50 ms, VENC 100 cm/s) was acquired at 1.5T using a 32-channel cardiac coil and SENSE=2. The 4D PC-MR velocity measurements were interpolated to the PIV resolution, and compared to PIV pixel-by-pixel after semi-automatic registration.
Results
Stroke volumes ranged from 12-37 ml and peak nozzle velocities from 19-40 cm/s. PIV velocities showed excellent agreement between days, both for the pixel-wise comparison (Figure 2A, R2=0.99, bias -0.04±0.72 cm/s) and for peak velocities (Figure 2B, R2=0.99, bias -1±2%). Validation of 4D PC-MR against PIV showed good pixel-wise agreement (Figure 2C, R2 = 0.95, bias 0.17±2.31 cm/s), but peak velocities were underestimated by 8-25% (Figure 2D, R2=0.89, bias -14±7%).
Conclusions
This study shows that 1) the proposed vortex ring flow phantom can be used for validation of 4D PC-MR measurements, 2) 4D PC-MR shows good overall agreement in velocity for the pixel-wise comparison, but underestimates peak velocities by 8-25% compared to PIV.
Funding
This study was supported by Swedish Research Council grants VR 621-2005-3129, VR 621-2008-2949, and VR K2009-65X-14599-07-3, National Visualization Program and Knowledge Foundation grant 2009-0080, the Medical Faculty at Lund University, Sweden, the Region of Scania, Sweden and the Swedish Heart-Lung Foundation.
References
Elkins CJ, Alley MT, Saetran L, Eaton JK: Three-dimensional magnetic resonance velocimetry measurements of turbulence quantities in complex flow. Exp Fluids. 2008, 46: 285-296.
Nilsson A, Bloch KM, Töger J, Heiberg E, Stahlberg F: Accuracy of four-dimensional phase-contrast velocity mapping for blood flow visualizations: a phantom study. Acta radiol. 2013
Knobloch V, Binter C, Gülan U, Sigfridsson A, Holzner M, Lüthi B, Kozerke S: Mapping mean and fluctuating velocities by Bayesian multipoint MR velocity encoding-validation against 3D particle tracking velocimetry. Magn Reson Med. 2014, 71: 1405-15. 10.1002/mrm.24785.
Liess C: Experimentelle Untersuchung Des Lebenslaufes von Ringwirbeln. 1978, Göttningen: Max-Planck-Institut für Strömungsforschung
Author information
Authors and Affiliations
Rights and permissions
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
About this article
Cite this article
Töger, J., Bidhult, S.L., Revstedt, J. et al. Phantom validation of 4D flow: independent validation of flow velocity quantification using particle imaging velocimetry. J Cardiovasc Magn Reson 17 (Suppl 1), O18 (2015). https://doi.org/10.1186/1532-429X-17-S1-O18
Published:
DOI: https://doi.org/10.1186/1532-429X-17-S1-O18