Gradient-induced voltages on 12-lead ECGs during high-duty-cycle MRI sequences and a theoretically-based method to remove them
Journal of Cardiovascular Magnetic Resonance volume 17, Article number: P243 (2015)
Background
An MRI-compatible 12-lead ECG platform, equipped with MRI-gradient induced-voltage removal hardware and magneto-hydrodynamic voltage removal software [1], was previously applied to physiological monitoring and synchronization of cardiac imaging of patients inside MRI. This approach had limited success for high-duty-cycle [(total Gradient-ramp-time per R-R)/(R-R time) >20%] MRI sequences, such as Steady State Free Precession (SSFP), Short-TR Gradient Echo (GRE), and Short-TR Fast Spin Echo. The study objective is to measure and develop a method to remove gradient-induced voltages on 12-lead ECGs during high-duty-cycle MRI sequences.
Methods
A modification of the hardware developed in [1] enabled measuring the gradient-induced voltage over a 24 kHz frequency-range and +/-10V, together with the x, y, and z gradient waveforms. ECGs were measured in 9 volunteers at 3T (Siemens Skyra). A theoretical equation for the gradient-induced voltages on each ECG electrode (Vi, where i=1,2,3..9 was derived, based on Maxwell's equations [2] and concomitant fields [3] [Fig. 1A]. It includes 1st and 2nd order gradient waveform terms and estimates induced voltages even on ECG electrodes positioned farthest from magnet iso-center, such as the limb leads.
The 19 equation coefficients were obtained during 6-8 second training sequences, consisting of highly accelerated (3-4sec, GRAPPA=6-8) versions of each sequence, followed by non-imaging segments. The non-imaging segments were used to obtain the shape of the true ECG traces over the entire R-R cycle ("template"). This template was subtracted from ECG traces acquired during imaging, resulting in the net gradient-induced voltages, which was then fit to the equation, providing the 19 coefficients for each electrode.
Multi-slice imaging was then performed, with real-time subtraction of the gradient-induced voltages from each acquired ECG trace, utilizing the computed Vi.
Results
Measured limb-lead ECG voltages during SSFP imaging, with the heart at iso-center (Fig. 1B), were 0.7-1.0 Volt PTP, with frequency components up to 20 KHz (Fig. 1C). Applying the equation for gradient-induced voltage removal during GRE (Fig. 1D-E), and multi-slice SSFP imaging (Fig. 1F), resulted in a normalized-difference between the recovered ECG trace and the true ECG of <20%. Equation coefficients varied by subject, sequence, sequence parameters, and slice orientation (Fig. 1G).
Conclusions
An equation was derived for the strong gradient-induced voltages observed in 12-lead ECGs during high-duty-cycle MRI sequences. A rapid training sequence permitted computing equation-coefficients, followed by real-time gradient-induced voltage removal during imaging.
Funding
NIH U41-RR019703, R03-EB013873-01A1, AHA 10SDG261039.
References
Tse : MRM. '13.
Bowtell : MRM. ‘00.
Bernstein : MRM. '89.
Author information
Authors and Affiliations
Rights and permissions
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
About this article
Cite this article
Zhang, S.H., Tse, Z.T., Dumoulin, C.L. et al. Gradient-induced voltages on 12-lead ECGs during high-duty-cycle MRI sequences and a theoretically-based method to remove them. J Cardiovasc Magn Reson 17 (Suppl 1), P243 (2015). https://doi.org/10.1186/1532-429X-17-S1-P243
Published:
DOI: https://doi.org/10.1186/1532-429X-17-S1-P243