Adjustment of the velocity encoding parameter to the blood flow velocity is not necessary for accurate and precise quantification of aortic regurgitation severity with phase contrast magnetic resonance imaging

Background

Accurate and precise quantification of aortic regurgitation (AVR) severity by cardiovascular magnetic resonance is essential for the clinical decission-making and timing of surgery. The regurgitant flow volume (RV) can be measured directly by 2D phase contrast (PC) velocity measurements. The velocity encoding parameter, venc, has been identified by others as an important factor for accurate and precise determination of RV. For large vessel and high signal to noise measurements, though, integration of the measured blood flow velocities over the vessel lumen and over the cardiac phases for calculation of RV should average out variations in the measured velocities and enable high precision estimates of RV independently of venc. Furthermore, application of a correction method that effectively reduces the background velocity offset in the PC image to a sufficiently low value should enable accurate estimation of RV and remove the venc dependency. The aim of the study was to demonstrate venc insensitivity in the estimated RV with effectively background offset corrected PC velocity measurements.

Methods

Measurements with high and low venc (~150 and 50 cm/s, respectively) were performed at the sinotubular junction on patients (n=28; 27-83y) and volunteers (n=26; 24-58y) using a 1.5 T scanner. Corrections for background offsets were automatically performed by the scanner and post-acquisition by means of adaptive image filtering. The mean background offset, standard deviation (sd) and coefficient of variation (cv) of repeated measurements were determined for the whole cohort. The mean RV, and sd and cv of RV were determined for patients and volunteers separately. For comparison, Wilcoxon signed-rank test was performed at a significance level of p<0.05.

Results

A background offset was present in all measurements. After correction, the offset was small (Fig 1) and did not cause any significant differences in RV regarding venc (p>0.8; Fig 2). Also, the variation in RV, determined from repeated measurements, did not significantly depend on venc (p>0.8).

Figure 1
figure 1

The background velocity offset after correction determined in ROI2 (), ROI3 (□), ROI4 (Δ) and ROI5 (×) in stationary muscle tissue for the high (AoH) and low venc measurement (AoL). Error bars in the mean show the error propagated sd from repeated measurements of individual subjects and error bars in cv show the sd among all subjects.

Figure 2
figure 2

The RV in volunteers (right) and patients (left) after correction for the high (AoH) and low venc measurement (AoL). Grey dots show the mean of repeated measurements of each subject and black dots show the mean of all subjects. Error bars, shown as the sd of the mean propagated from repeated measurements in individual subjects, was at some points smaller than the symbols.

Conclusions

We have demonstrated venc insensitivity in both the accuracy and precision of RV using background offset corrected PC velocity measurements. Without the need for adjustment of venc to the blood flow velocity, the time for the examination will be substantially reduced.

Funding

This study was funded by a project grant from the Health & Medical Care Committee of the Regional Executive Board, Västra Götaland Region, Sweden.

Author information

Authors and Affiliations

Authors

Rights and permissions

Open Access  This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made.

The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder.

To view a copy of this licence, visit https://creativecommons.org/licenses/by/4.0/.

The Creative Commons Public Domain Dedication waiver (https://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Svensson, F., Polte, C.L., Johnsson, Å.A. et al. Adjustment of the velocity encoding parameter to the blood flow velocity is not necessary for accurate and precise quantification of aortic regurgitation severity with phase contrast magnetic resonance imaging. J Cardiovasc Magn Reson 17 (Suppl 1), P329 (2015). https://doi.org/10.1186/1532-429X-17-S1-P329

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/1532-429X-17-S1-P329

Keywords