Joint denoising and motion correction: initial application in single-shot cardiac MRI

Background

Single-shot (SSH) pulse sequences in CMR are beneficial for rapid image acquisition that is robust to motion, especially in arrhythmic patients or poor breath-holders. However, this fast scanning technique trades scan time for a lower signal-to-noise ratio compared to conventional multi-shot acquisitions. Here we propose a motion-compensated denoising technique that improves the image quality from multiple free-breathing single-shot acquisitions.

Methods

Acquisition - In 7 patients with suspected cardiovascular disease, 4 repetitions of SSH Late Gadolinium Enhancement (LGE) were acquired in free-breathing 7-10 minutes after Gd injection [1]. The experiments were performed on 3T MR750w and 1.5T MR450w systems (GE Healthcare, WI, USA) with the following parameters: 8-mm slice, acceleration factor=2. The scan time of SSH LGE imaging (8s/slice in free-breathing) was shorter than the multi-shot LGE (13s/slice in breath-hold).

Reconstruction - The proposed method consists of 3 steps: 1) SENSE reconstruction of the undersampled data, 2) extraction of the motion that occurred between the repetitions with a non-rigid registration, and 3) motion compensated joint denoising of the whole dataset with Beltrami regularization, which offers an ideal compromise between feature preservation and staircasing reduction [2]. Steps 2 and 3 are applied iteratively. The results were compared to the non-iterative previously proposed method [3] where the third step is replaced by a simple average of the registered images.

Results

Example reconstructions results are shown in Fig 1. Images reconstructed by employing a standard averaging of 4 SSH LGE images acquired during free-breathing show a result spatially blurred and noisy (b). The benefit of motion correction is seen on (c), and the benefit of denoising in (d). Compared to the non-iterative method, one can see that the proposed joint denoising and motion correction method reconstructs sharper edges (arrows) while reducing the noise.

Figure 1
figure 1

Cardiac imaging reconstructions after SSH LGE acquisition in free breathing. a) One Repetition (192x152), b) Average of 4 Repetitions, c) Iterative motion-compensated reconstruction with averaging instead of denoising, d) Iterative Reconstruction with Beltrami regularization. e) Conventional multi-shot LGE in breath-hold (224x192).

Conclusions

We introduced a new joint denoising and motion correction algorithm designed for fast, robust cardiac imaging. By iteratively incorporating the estimated motion into the reconstruction process, we increase the robustness of the model and exhibit good quality imaging. A limitation to the method is that potential through-plane motion cannot be corrected, although it remains negligible compared to the slice thickness. Ultimately, this method could be beneficial to both cardiac exam workflow and patient comfort since it allows robust image quality without the need for breath-holds.

Funding

This work was supported by the European Commission, through Grant 605162.

References

  1. Schelbert et al: Circ Cardiovasc Imaging. 2010

    Google Scholar 

  2. Zosso et al: UCLA Cam Report. 2014

    Google Scholar 

  3. Kellman et al: MRM. 2005

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Bustin, A., Janich, M.A., Brau, A.C. et al. Joint denoising and motion correction: initial application in single-shot cardiac MRI. J Cardiovasc Magn Reson 17 (Suppl 1), Q29 (2015). https://doi.org/10.1186/1532-429X-17-S1-Q29

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/1532-429X-17-S1-Q29

Keywords